Purpose: A fully independent, machine learning-based automatic treatment couch parameters prediction was developed to support surface guided radiation therapy (SGRT)-based patient positioning protocols. Additionally, this approach also acts as a quality assurance tool for patient positioning.

Materials/methods: Setup data of 183 patients, divided into four different groups based on used setup devices, was used to calculate the difference between the predicted and the acquired treatment couch value.

Results: Couch parameters can be predicted with high precision . A significant difference (p < 0.01) between the variances of Lung and Brain patients was found. Outliers were not related to the prediction accuracy, but are due to inconsistencies during initial patient setup.

Conclusion: Couch parameters can be predicted with high accuracy and can be used as starting point for SGRT-based patient positioning. In case of large deviations (>1.5 cm), patient setup has to be verified to optimally use the surface scanning system.