Aims: To evaluate safety of leadless pacemaker implantation through the internal jugular vein in a larger cohort with longer follow-up. Moreover, feasibility of non-apical pacing as well as relation between pacing site and QRS duration were assessed.

Methods: Eighty Two consecutive patients, who received a leadless pacemaker though the internal jugular vein, were included. Electrical parameters were measured at regular follow-up and any complications were registered. Paced QRS interval was compared for three pacing sites, RVOT, RV mid septum, and RV apical septum.

Results: In all patients, the leadless pacemaker was implanted successfully. In 69 patients, the device was implanted in a non-apical position. In 71% of cases, the device could be deployed at first attempt. The median fluoroscopy time was 4.4 min (range 0.9-51) The paced QRS interval was significantly narrower for non-apical pacing sites compared to apical pacing si 156  vs. 179 ms. p = .04, respectively. During mean follow-up of 16 months (range 0-43 months), electrical parameters remained stable. Two complications occurred, which could be resolved during the implant procedure. There were no access site related complications.

Conclusion: The jugular approach for leadless pacemaker implantation is feasible and may avoid vascular complications. It facilitates non-apical positioning of leadless pacemakers leading to a narrower paced QRS interval. The jugular approach allows for immediate post procedural ambulation.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.14587DOI Listing

Publication Analysis

Top Keywords

leadless pacemaker
16
jugular approach
12
paced qrs
12
qrs interval
12
approach leadless
8
pacemaker implantation
8
internal jugular
8
jugular vein
8
non-apical pacing
8
electrical parameters
8

Similar Publications

Introduction: The advancement of medical technology has introduced leadless pacemakers (LPMs) as a significant innovation in cardiac pacing, offering potential advantages over traditional ventricular transvenous pacemakers. This report explores the application of LPMs in two patients with complex valvular histories, particularly those with mechanical tricuspid valves.

Case Reports: The first case involves a 60-year-old male with a history of rheumatic heart disease and triple valve replacement who developed a high-grade AV block.

View Article and Find Full Text PDF

Background: Achieving hemostasis of large bore venous access sites can be challenging and time consuming. Closure devices have proven to be superior in achieving hemostasis, reducing time to ambulation and improving patient comfort, compared to manual hemostasis techniques after femoral venous and arterial access. The closure of the jugular vein following large bore access has not been investigated in previous studies.

View Article and Find Full Text PDF

A rare cardiac presentation of a lymphoma: case report.

Eur Heart J Case Rep

January 2025

Department of Cardiology, Klinik Floridsdorf, Brünnerstraße 68, Vienna 1210, Austria.

Background: Cardiac lymphoma is a rare disease that can present in various ways. Additionally, atypical clinical presentation makes the diagnosis even more challenging. The most common type of cardiac lymphoma is diffuse large B-cell lymphoma.

View Article and Find Full Text PDF

Minimizing for Maximum Benefit: An Illustrative Case-Series of Atrial Only Leadless Pacing.

Pacing Clin Electrophysiol

December 2024

Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York Presbyterian Hospital, New York, USA.

Leadless pacing technology now includesdedicated atrial helix-fixation leadless pacemakers (LPs), expanding theapplication of leadless devices for patients with sinus node dysfunction andatrioventricular block during sinus rhythm. This first reportedcase-series of atrial LPs describes and discusses the potential use-casescenarios of recently approved helix-fixation atrial LPs. The article highlights important concepts regarding their use, including implantationtechniques, programming, battery conservation, and the low rate of progressionof AV block in patients implanted with AAI(R) pacemakers.

View Article and Find Full Text PDF

Reel's syndrome (RS) is an unusual cause of pacemaker lead dislodgement. We present the case of a 59-year-old female patient with Down syndrome (DS) implanted with a dual-chamber endovascular pacemaker due to symptomatic sinus node disfunction, reporting several syncopal episodes in last days and showing abnormal electrical parameters at the 2-months follow-up due to RS. The malfunctioning device was removed and an endocardial leadless pacing system was implanted.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!