Repeatability of new optical biometer and agreement with 2 validated optical biometers, all based on SS-OCT.

J Cataract Refract Surg

From the Studio Oculistico d'Azeglio, Bologna, Italy (Galzignato, Barboni); University Eye Clinic, S. Orsola-Malpighi Hospital, University of Bologna, Bologna, Italy (Lupardi); Stein Eye Institute, University of California, Los Angeles, California (Hoffer); St. Mary's Eye Center, Santa Monica, California (Hoffer); San Raffaele Hospital, Milan, Italy (Barboni); IRCCS G.B. Bietti Foundation, Rome, Italy (Schiano-Lomoriello, Savini).

Published: January 2023

Purpose: To evaluate the repeatability of the measurements provided by a new optical biometer (EyeStar 900) based on swept-source optical coherence tomography (SS-OCT) and their agreement with the measurements given by 2 validated biometers based on the same technology, the IOLMaster 700 and Argos.

Setting: IRCCS G.B. Bietti Foundation, Rome, Italy.

Design: Prospective evaluation of diagnostic test.

Methods: In a series of unoperated eyes, 3 consecutive scans were acquired with the EyeStar 900, and 1 with the IOLMaster 700 and the Argos. The following biometry parameters were analyzed: axial length (AL), keratometry (K), corneal astigmatism, central corneal thickness, corneal diameter (CD), anterior chamber depth (ACD), lens thickness (LT), and lens tilting. Repeatability was assessed using test-retest variability, the coefficient of variation (CoV), and the intraclass correlation coefficient (ICC); agreement was based on the 95% limits of agreement.

Results: 56 eyes of 56 patients were analyzed. High repeatability was achieved for all measured parameters, as the CoV was <1% in most cases and ICC was >0.95 for all parameters. Good to high agreement was found among the measurements of the 3 optical biometers, although some statistically significant differences were detected between the EyeStar 900 and Argos (mean K, ACD, LT, and CD were higher with the Argos). The Argos measured a shorter AL in eyes >25 mm.

Conclusions: The new generation SS-OCT EyeStar 900 optical biometer produces highly repeatable measurements that are in good agreement with those provided by 2 previously validated instruments.

Download full-text PDF

Source
http://dx.doi.org/10.1097/j.jcrs.0000000000001023DOI Listing

Publication Analysis

Top Keywords

eyestar 900
16
optical biometer
12
optical biometers
8
biometers based
8
agreement measurements
8
iolmaster 700
8
agreement
5
optical
5
repeatability
4
repeatability optical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!