Background: When using an external ventricular drain (EVD) to monitor intracranial pressure (ICP), nurses need to know how long to wait after each manipulation of the transducer before the displayed ICP value represents an accurate signal. This study explores ICP signal equilibration time (EqT) under clinical conditions.

Methods: This was a prospective ex vivo study using a simulated skull, standard EVD tubing, and a strain gauge transducer. All 270 trials simulating 90 combinations of different pressures and common clinical conditions were completed in August 2021. Each trial was recorded on video. Videos were scored using video editing software to obtain the exact start and stop time for each trial.

Results: The mean EqT was 44.90 (18.77) seconds. One hundred fifty (55.56%) observations did not reach their expected value within 60 s. The longest mean EqTs were noted when blood was present in the EVD tubing (57.67 [8.91] seconds), when air bubbles were in the tubing (57.41 [8.73] seconds), and when EVD tubing was not flat (level) (50.77 [15.43] seconds). An omnibus test comparing mean EqT for conditions with no variables manipulated (30.08 [16.07] seconds) against mean EqT for all others (47.18 [18.13] seconds) found that mean EqTs were significantly different (P < 0.001).

Conclusions: Even when no additional variables were introduced, the mean EqTs were ~ 30 s. Common clinical variables increase the length of time before a transducer connected to an EVD will provide an accurate reading. Nurses should wait at least 30 s after turning the EVD stopcock before assuming ICP value reflects accurate ICP.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12028-022-01581-5DOI Listing

Publication Analysis

Top Keywords

evd tubing
12
intracranial pressure
8
seconds
6
predictive values
4
values time
4
time transducer
4
transducer stopcock
4
stopcock closure
4
closure accurate
4
accurate intracranial
4

Similar Publications

Treatment of post-thalamic hemorrhage hydrocephalus: ventriculoperitoneal shunt or endoscopic third ventriculostomy? A retrospective observational study.

BMC Neurol

September 2024

Departmen of Rehabilitation Medicine Center, People's Hospital of Ningxia Hui Autonomous Region, Ningxia Hui Autonomous Region, No. 301 Zhengyuan North Street, Jinfeng District, Yinchuan, 750002, China.

Background: The aim of this study was to compare the efficacy of ventriculoperitoneal shunt (VPS) and endoscopic third ventriculostomy (ETV) for the treatment of hydrocephalus after thalamic hemorrhage (TH) where external ventricular drainage (EVD) could not be removed after hematoma absorption, and to provide a theoretical basis for the clinical treatment of hydrocephalus after TH.

Methods: The clinical data of patients with hydrocephalus after TH whose EVD could not be removed after hematoma absorption were retrospectively analyzed. According to the patients' surgical methods, the patients were divided into the VPS group and ETV group.

View Article and Find Full Text PDF

Purpose: To develop and validate a noninvasive imaging technique for accurately assessing very slow CSF flow within shunt tubes in pediatric patients with hydrocephalus, aiming to identify obstructions that might impede CSF drainage.

Theory And Methods: A simulation of shunt flow enhancement of signal intensity (shunt-FENSI) signal is used to establish the relationship between signal change and flow rate. The quantification of flow enhancement of signal intensity data involves normalization, curve fitting, and calibration to match simulated data.

View Article and Find Full Text PDF

Objective: Abdominal CSF pseudocysts are an uncommon but challenging complication of ventriculoperitoneal shunts. Pseudocysts consist of a loculated intraperitoneal compartment that inadequately absorbs CSF and may be infected or sterile at diagnosis. The treatment goal is to clear infection if present, reduce inflammation, and reestablish long-term function in an absorptive (intraperitoneal) space.

View Article and Find Full Text PDF

Background: When using an external ventricular drain (EVD) to monitor intracranial pressure (ICP), nurses need to know how long to wait after each manipulation of the transducer before the displayed ICP value represents an accurate signal. This study explores ICP signal equilibration time (EqT) under clinical conditions.

Methods: This was a prospective ex vivo study using a simulated skull, standard EVD tubing, and a strain gauge transducer.

View Article and Find Full Text PDF

Insertion of an external ventricular drain (EVD) is one of the most common and most important lifesaving procedures encountered in the neurologic intensive care unit, but often done by the junior members of the team. A good number of complications may follow the insertion of EVD. In the available literature, only one case was reported with the placement of EVD in suprasellar cistern.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!