Purpose: Cranial stereotactic radiotherapy (SRT) requires highly accurate lesion delineation. However, MRI can have significant inherent geometric distortions. We investigated how well the Elements Cranial Distortion Correction algorithm of Brainlab (Munich, Germany) corrects the distortions in MR image-sets of a phantom and patients.
Methods: A non-distorted reference computed tomography image-set of a CIRS Model 603-GS (CIRS, Norfolk, VA, USA) phantom was acquired. Three-dimensional T1-weighted images were acquired with five MRI scanners and reconstructed with vendor-derived distortion correction. Some were reconstructed without correction to generate heavily distorted image-sets. All MR image-sets were corrected with the Brainlab algorithm relative to the computed tomography acquisition. CIRS Distortion Check software measured the distortion in each image-set. For all uncorrected and corrected image-sets, the control points that exceeded the 0.5-mm clinically relevant distortion threshold and the distortion maximum, mean, and standard deviation were recorded. Empirical cumulative distribution functions (eCDF) were plotted. Intraclass correlation coefficient (ICC) was calculated. The algorithm was evaluated with 10 brain metastases using Dice similarity coefficients (DSC).
Results: The algorithm significantly reduced mean and standard deviation distortion in all image-sets. It reduced the maximum distortion in the heavily distorted image-sets from 2.072 to 1.059 mm and the control points with > 0.5-mm distortion fell from 50.2% to 4.0%. Before and especially after correction, the eCDFs of the four repeats were visually similar. ICC was 0.812 (excellent-good agreement). The algorithm increased the DSCs for all patients and image-sets.
Conclusion: The Brainlab algorithm significantly and reproducibly ameliorated MRI distortion, even with heavily distorted images. Thus, it increases the accuracy of cranial SRT lesion delineation. After further testing, this tool may be suitable for SRT of small lesions.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00066-022-01988-1 | DOI Listing |
Phys Eng Sci Med
January 2025
Physics Department, Instituto Zunino, Obispo Oro 423, X5000BFI, Córdoba, Argentina.
Treatment Planning Systems (TPS) offer algorithms for distortion correction (DC) of Magnetic Resonance (MR) images, whose performances demand proper evaluation. This work develops a procedure using a virtual phantom to quantitatively assess a TPS DC algorithm. Variations of the digital Brainweb MR study were created by introducing known distortions and Control Points (CPs).
View Article and Find Full Text PDFNeuroimage
January 2025
Key Laboratory of Ultra-Weak Magnetic Field Measurement Technology, Ministry of Education, School of Instrumentation and Optoelectronic Engineering, Beihang University, 100191, Beijing, China; Hangzhou Institute of Extremely-Weak Magnetic Field Major National Science and Technology Infrastructure, Hangzhou, 310051, China; State Key Laboratory of Traditional Chinese Medicine Syndrome/Health Construction Center, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, 510120, China; Hefei National Laboratory, Hefei, 230088, China. Electronic address:
Optically pumped magnetometer-based magnetoencephalography (OPM-MEG) is an novel non-invasive functional imaging technique that features more flexible sensor configurations and wearability; however, this also increases the requirement for environmental noise suppression. Subspace projection algorithms are widely used in MEG to suppress noise. However, in OPM-MEG systems with a limited number of channels, subspace projection methods that rely on spatial oversampling exhibit reduced performance.
View Article and Find Full Text PDFSci Rep
January 2025
Department of Electronic Engineering, Hanyang University, Seoul, 04763, South Korea.
This paper presents novel MIMO microstrip patch antennas with dimensions of 40 × 80 × 1.6 mm³ incorporating a decoupling and pattern correction structure (DPCS) designed to mitigate mutual coupling and radiation pattern distortion, operating within 3.6-3.
View Article and Find Full Text PDFJ Med Imaging (Bellingham)
December 2024
University of Houston, Department of Physics, Houston, Texas, United States.
Purpose: Photon counting detectors offer promising advancements in computed tomography (CT) imaging by enabling the quantification and three-dimensional imaging of contrast agents and tissue types through simultaneous multi-energy projections from broad X-ray spectra. However, the accuracy of these decomposition methods hinges on precise composite spectral attenuation values that one must reconstruct from spectral micro-CT. Errors in such estimations could be due to effects such as beam hardening, object scatter, or detector sensor-related spectral distortions such as fluorescence.
View Article and Find Full Text PDFWorld J Clin Cases
December 2024
Department of Ophthalmology, Tung Wah Eastern Hospital, Hong Kong 999077, China.
Upper lid blepharoplasty is a surgical procedure that requires meticulous planning and execution. It is not a standalone procedure but can be combined with ptosis correction surgery. Prior to surgery, thorough lid examinations are essential to determine the appropriate approaches.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!