Background: Arterial hypertension is a worldwide public health problem. While it is currently diagnosed and monitored non-invasively using the oscillometric method, having the ability to measure blood pressure (BP) using a smartphone application could provide more widespread access to hypertension screening and monitoring. In this observational study in intensive care unit patients, we compared blood pressure values obtained using a new optical smartphone application (OptiBP™; test method) with arterial BP values obtained using a radial artery catheter (reference method) in order to help validate the technology.

Methods: We simultaneously measured three BP values every hour for five consecutive hours on two consecutive days using both the smartphone and arterial methods. Bland-Altman and error grid analyses were used for agreement analysis between both approaches. The performance of the smartphone application was investigated using the Association for the Advancement of Medical Instrumentation (AAMI) and the International Organization for Standardization (ISO) definitions, which require the bias ± SD between two technologies to be below 5 ± 8 mmHg.

Results: Among the 30 recruited patients, 22 patients had adequate OptiBP™ values and were thus analyzed. In the other 8 patients, no BP could be measured due to inadequate signals. The Bland-Altman analysis revealed a mean of the differences ± SD between both methods of 0.9 ± 7 mmHg for mean arterial pressure (MAP), 0.2 ± 14 mmHg for systolic arterial pressure (SAP), and 1.1 ± 6 mmHg for diastolic arterial pressure (DAP). Error grid analysis demonstrated that the proportions of measurement pairs in risk zones A to E were 88.8% (no risk), 10% (low risk), 1% (moderate risk), 0% (significant risk), and 0% (dangerous risk) for MAP and 88.4%, 8.6%, 3%, 0%, 0%, respectively, for SAP.

Conclusions: This method comparison study revealed good agreement between BP values obtained using the OptiBP™ and those done invasively. The OptiBP™ fulfills the AAMI/ISO universal standards for MAP and DAP (but not SAP). Error grid showed that the most measurements (≥ 97%) were in risk zones A and B.

Trial Registration: ClinicalTrials.gov registration: NCT04728477.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9376576PMC
http://dx.doi.org/10.1186/s12871-022-01797-0DOI Listing

Publication Analysis

Top Keywords

blood pressure
16
smartphone application
12
error grid
12
arterial pressure
12
optical smartphone
8
method comparison
8
comparison study
8
intensive care
8
care unit
8
unit patients
8

Similar Publications

In this study we have used a highly immersive virtual reality (VR) cycling environment where incongruence between virtual hill gradient (created by visual gradient and bike tilt angle) and actual workload (pedalling resistance) can experimentally manipulate perception of exercise effort. This therefore may provide a method to examine the role of effort perception in cardiorespiratory control during exercise. Twelve healthy untrained participants (7 men, age 26 ± 5 years) were studied during five visits.

View Article and Find Full Text PDF

Background: There are limited therapeutic options in cases of failed reperfusion (modified thrombolysis in cerebral infarction [mTICI] score < 2b) after stent-retriever and/or aspiration based endovascular treatment (EVT) for acute ischemic stroke. Despite the absence of data supporting its use, rescue therapy (balloon angioplasty and/or stent implantation) is often utilized in such cases. Studies are limited to large vessel occlusions, while the outcomes and complications after rescue therapy in medium/distal vessel occlusions (MDVOs) have not been reported.

View Article and Find Full Text PDF

We investigated the effects of low and high frequency isometric handgrip exercise training (IHGT) on resting blood pressure, and the affective/perceptual responses during training. Sixty young normotensive adults were randomised to either a no-intervention control group (CON:  = 20; 12 female) or a group performing either two (LOW:  = 20; 18 female) or four (HIGH:  = 20; 13 female) sessions/week of IHGT for 4 weeks. IHGT involved 4 × 2-min holds at 30% maximal voluntary contraction using the dominant hand.

View Article and Find Full Text PDF

Objective: ADHD is one of the most common neurodevelopmental disorders, seen in children and adolescents, and is often treated with various pharmacological agents, especially methylphenidate. There are differing opinions in the literature regarding the cardiovascular safety of long-term methylphenidate use. Studies suggest that the drug may increase the risk of hypertension, myocardial infarction, ventricular arrhythmia, sudden cardiac death, cardiomyopathy, heart failure (HF), pulmonary hypertension, and stroke.

View Article and Find Full Text PDF

Atherosclerotic cardiovascular disease (ASCVD) risk calculators estimate the 10-year incident risk of myocardial infarction (MI), coronary artery disease (CAD) death, or stroke; however, they lack comprehensiveness and accuracy. Carotid intima-media thickness (CIMT) is a surrogate marker that may improve risk estimation acumen. The objective of this study was to derive ASCVD risk scores from historical data and determine whether these risk scores are associated with the history of subclinical CAD and CIMT.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!