A 72-year-old male with severe dilated cardiomyopathy, benefited of implantable cardiac defibrillator implementation. Device control shows high impedance. On X-ray, electrodes were completely twisted in the generator pocket, they were replaced and the generator was fixed to pectoralis-major fascia. Nurses report patient abnormal movements, scratching implantation area. This was identified as probably the trigger of the complication.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9372525 | PMC |
http://dx.doi.org/10.1016/j.radcr.2022.07.035 | DOI Listing |