Purpose: To evaluate microscopically whether the print quality and accuracy of sizing of Landolt ring near vision charts are adequate for the calibration of reading charts.
Methods: Near vision charts with Landolt rings from Oculus GmbH (C-Test; Wetzlar, Germany), Precision Vision (Woodstock, IL) and the RADNER Charts were examined, as well as custom-made Landolt rings optimized for print quality. Microscopic investigations and measurements were performed by using a Huvitz HSZ 600 stereomicroscope (Nikon NIS Elements software) to evaluate the height of the Landolt rings, the thickness of the lines, and the width of the openings. The deviations from the mathematically correct values, which were calculated as given in the EN/ISO 8596 and by the International Council of Ophthalmology (ICO), were analyzed (calculated for a test distance of 40 cm).
Results: All the near vision charts showed notable deficiencies in print quality and aberrations from the nominal values in the height, thickness of the lines, and width of the openings. The openings were too narrow, whereas the height and thickness of the lines were larger than the nominal values. Even the openings of Landolt rings optimized for print quality were not always within an acceptable 5% tolerance and need further improvement.
Conclusion: This study reports inaccuracies in the heights, thicknesses of the lines, and widths of the openings of Landolt rings in all the near vision charts investigated. The extent of these inaccuracies excludes such near vision charts as reference tests for the calibration of reading charts. The x-height in relation to the visual angle still seems to be the most reliable method for standardizing the print sizes for reading charts.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9377069 | PMC |
http://dx.doi.org/10.1186/s40662-022-00302-5 | DOI Listing |
Alzheimers Dement
December 2024
Centre for Brain Research, Indian Institute of Science, Bangalore, Karnataka, India.
Background: Refractive errors are common visual comorbidities among the elderly. Cognitive dysfunction also occurs in this population. A study by Ong et al (2013) demonstrated an association of refractive errors with poor cognitive performance.
View Article and Find Full Text PDFJ Pediatr Ophthalmol Strabismus
December 2024
Purpose: To investigate whether wide-angle fluorescein angiography (FA) was useful for detecting familial exudative retinopathy (FEVR) in patients with neurodevelopmental disorder with spastic diplegia and visual defects (NEDSDV).
Methods: This was a retrospective chart review from 2013 to 2023 of all consecutive patients with NEDSDV and pathogenic or likely pathogenic variants.
Results: Seven patients (four females, three males) were seen in the ophthalmology clinic (median age: 14 months).
J West Afr Coll Surg
August 2024
Department of Ophthalmology, Federal Teaching Hospital Owerri, Imo State, Nigeria.
Background: Colour vision defect (CVD) is a public health issue with approximately one in ten males exhibiting some form of colour deficiency. Colour is used extensively in education so CVD has a dramatic impact on the learning, educational and social potentials of children. Racial differences have been reported with higher values noted amongst people of European ancestry.
View Article and Find Full Text PDFRes Rep Trop Med
December 2024
Global Health Institute, University of Antwerp, Antwerp, Belgium.
Introduction: Raga County is an onchocerciasis-endemic area in the Western Bahr El Ghazal state of South Sudan, known to have a high prevalence of blindness. The objective of this study was to determine the causes of eye disease and blindness in Raga County as well as to assess the relationship of eye diseases with other prevalent conditions like onchocerciasis and epilepsy.
Methods: We reviewed unpublished pre-community directed treatment with ivermectin (CDTI) data about eye disease and onchocerciasis in Western Bahr El Ghazal including Raga.
Ophthalmic Plast Reconstr Surg
December 2024
Hariram Motumal Nasta & Renu Hariram Nasta Ophthalmic Plastic Surgery Services, KAR Campus.
Purpose: To evaluate the growth, management, and outcomes of epibulbar dermolipomas over a 5-year follow-up period.
Methods: This was a retrospective chart review of epibulbar dermolipoma patients with a minimum follow-up of 5 years, which analyzed the changes in size, refractive errors (spherical equivalent), best-corrected visual acuity, histology, and surgical outcomes.
Results: A total of 61 eyes of 53 patients (32 females) with an average presenting age of 4.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!