Background: Methods for accurate absolute dose (AD) calibration are essential for the proper functioning of radiotherapy treatment machines. Many systems do not conform to TG-51 calibration standards, and modifications are required. TG-21 calibration is also a viable methodology for these situations with the appropriate setup, equipment, and factors. It has been shown that both these methods result in minimal errors. A similar approach has been taken in calibrating the dose for a recent vault-free radiosurgery system.

Purpose: To evaluate modified TG-21 and TG-51 protocols for AD calibrations of the ZAP-X radiosurgery system using ion chambers, film, and thermoluminescent dosimeters (TLDs).

Methods: The current treatment planning system for ZAP-X requires AD calibration at d (7 mm) and 450 mm source-to-axis distance. Both and N [R/C] calibration coefficients were provided by an accredited dosimetry calibration laboratory for a physikalisch technische werkstatten (PTW) 31010 chamber (0.125 cc). The vendor provides an f-bracket that can be mounted on the collimator. Various phantoms can then be attached to the f-bracket. A custom acrylic phantom was designed based on recommendations from TG-21 and technical report series-398 that places the chamber at 500 mm from the source with a depth of 44-mm acrylic and 456-mm SSD. N along with other TG-21 parameters was used to calculate the AD. Measurements using a PTW MP3-XS water tank and the same chamber were used to calculate AD using and TG-51 factors. Dose verification was performed using Gafchromic film and 3rd party TLDs.

Results: Measurements from TG-51, TG-21 (utilizing the custom acrylic phantom), film, and TLDs agreed to within ± 2%.

Conclusions: A modified TG-51 AD calculation in water is preferred but may not be practical due to the difficulty in tank setup. The TG-21 modified protocol using a custom acrylic phantom is an accurate alternative option for dose calibration. Both of these methods are within acceptable agreement and provide confidence in the system's AD calibration.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15912DOI Listing

Publication Analysis

Top Keywords

custom acrylic
12
acrylic phantom
12
absolute dose
8
vault-free radiosurgery
8
radiosurgery system
8
calibration
8
dose calibration
8
tg-21
6
dose
5
tg-51
5

Similar Publications

Statement Of Problem: The angled screw channel (ASC) design has been well accepted for implant prostheses. However, investigation into the behavior of the ASC connection is sparse.

Purpose: The purpose of this in vitro study was to assess the effect of cyclic loading on the internal connection of an ASC system compared with straight access systems by measuring reverse torque values (RTVs) and using microcomputed tomography (µCT) imaging.

View Article and Find Full Text PDF

Objectives To assess the influence of cigarette smoke (CS) on the color and surface roughness of 3D printed, milled, and traditionally fabricated provisional crown and bridge (PC&B) materials. Materials and methods 112 disc-shaped samples were made employing four techniques and materials (28 per group) to fabricate PC&B prostheses. Specimens were fabricated using standard protocols, such as 3D printing, milling, conventional bis-acrylic resin, and traditional autopolymerizing polymethyl methacrylate (PMMA) resin.

View Article and Find Full Text PDF

In this work, magnetic molecularly imprinted polymer (MMIP) capable of selectively recognizing and adsorbing cordycepin was prepared. The MMIP was prepared using cordycepin as the template molecule, methacrylic acid and acrylamide as the functional monomer and ethylene glycol dimethacrylate as the crosslinker. The MMIP was analyzed using various techniques including transmission electron microscopy, thermogravimetric analysis, Fourier transform infrared spectroscopy, vibrating sample magnetometer and x-ray diffraction.

View Article and Find Full Text PDF

Background: Cementoplasty has been successfully used for treating fractures in various parts of the human body, although the use in weight-bearing long bones is a subject of controversial debate. Strategies to improve the mechanical properties of polymethylmethacrylate-based bone cement (BC) comprise changing the chemical composition or the application of metal reinforcement strategies. In clinical practice reinforced bone cement is used despite biomechanical basic research regarding this topic being scare.

View Article and Find Full Text PDF

Objectives: This study aimed to investigate the effect of a customized low-sodium bioactive glass (BG) fillers in experimental resin composites on their bond strength to dentin after 12 months of artificial aging. Specifically, it evaluated whether the bond strength was affected by different BG concentrations (0, 10, 20, 40 wt%) and artificial aging durations (1, 6, and 12 months).

Materials And Methods: Experimental composites were prepared with 10, 20, and 40 wt% of a customized low-sodium fluoride-containing BG.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!