Background: In cone-beam computed tomography (CBCT)-guided radiotherapy, off-by-one vertebral-body misalignments are rare but serious errors that lead to wrong-site treatments.
Purpose: An automatic error detection algorithm was developed that uses a three-branch convolutional neural network error detection model (EDM) to detect off-by-one vertebral-body misalignments using planning computed tomography (CT) images and setup CBCT images.
Methods: Algorithm training and test data consisted of planning CTs and CBCTs from 480 patients undergoing radiotherapy treatment in the thoracic and abdominal regions at two radiotherapy clinics. The clinically applied registration was used to derive true-negative (no error) data. The setup and planning images were then misaligned by one vertebral-body in both the superior and inferior directions, simulating the most likely misalignment scenarios. For each of the aligned and misaligned 3D image pairs, 2D slice pairs were automatically extracted in each anatomical plane about a point within the vertebral column. The three slice pairs obtained were then inputted to the EDM that returned a probability of vertebral misalignment. One model (EDM ) was trained solely on data from institution 1. EDM was further trained using a lower learning rate on a dataset from institution 2 to produce a fine-tuned model, EDM . Another model, EDM , was trained from scratch using a training dataset composed of data from both institutions. These three models were validated on a randomly selected and unseen dataset composed of images from both institutions, for a total of 303 image pairs. The model performances were quantified using a receiver operating characteristic analysis. Due to the rarity of vertebral-body misalignments in the clinic, a minimum threshold value yielding a specificity of at least 99% was selected. Using this threshold, the sensitivity was calculated for each model, on each institution's test set separately.
Results: When applied to the combined test set, EDM , EDM , and EDM resulted in an area under curve of 99.5%, 99.4%, and 99.5%, respectively. EDM achieved a sensitivity of 96% and 88% on Institution 1 and Institution 2 test set, respectively. EDM obtained a sensitivity of 95% on each institution's test set. EDM achieved a sensitivity of 95% and 88% on Institution 1 and Institution 2 test set, respectively.
Conclusion: The proposed algorithm demonstrated accuracy in identifying off-by-one vertebral-body misalignments in CBCT-guided radiotherapy that was sufficiently high to allow for practical implementation. It was found that fine-tuning the model on a multi-facility dataset can further enhance the generalizability of the algorithm.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.15927 | DOI Listing |
Eur Arch Otorhinolaryngol
January 2025
Department of Otolaryngology-Head and Neck Surgery, College of Medicine, King Saud University, Riyadh, Saudi Arabia.
Objective: This study aims to compare the outcomes of balloon-assisted rib graft placement with traditional graft placement in Endoscopic Posterior Cricoid Split with Rib Graft Placement (EPCS/RG).
Methods: We conducted a retrospective analysis of 23 patients who underwent EPCS/RG by a single senior surgeon at King Saud University Medical City from 2017 to 2024. Data were collected on demographics, surgical approach, and operative time.
Eur Arch Otorhinolaryngol
January 2025
Motion Sickness and Human Performance Laboratory, The Israel Naval Medical Institute, IDF Medical Corps, Haifa, Israel.
Purpose: Acute acoustic trauma (AAT) is a sudden sensorineural hearing loss (SNHL) due to exposure to high intensity impulse noise. There are no acceptable treatment guidelines, although several studies showed steroids could be effective in restoring hearing levels. A recent report suggested that steroids combined with hyperbaric oxygen therapy (HBOT) are a superior regiment for AAT.
View Article and Find Full Text PDFNPJ Digit Med
January 2025
Neurofibromatosis Type 1 Center and Laboratory for Neurofibromatosis Type 1 Research, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200011, China.
Deep-learning models have shown promise in differentiating between benign and malignant lesions. Previous studies have primarily focused on specific anatomical regions, overlooking tumors occurring throughout the body with highly heterogeneous whole-body backgrounds. Using neurofibromatosis type 1 (NF1) as an example, this study developed highly accurate MRI-based deep-learning models for the early automated screening of malignant peripheral nerve sheath tumors (MPNSTs) against complex whole-body background.
View Article and Find Full Text PDFAm J Kidney Dis
January 2025
Department of Medical Social Sciences, Northwestern University Feinberg School of Medicine, Chicago, USA; Northwestern University Transplant Outcomes Research Collaborative, Comprehensive Transplant Center, Feinberg School of Medicine, Chicago, IL, USA; Centre for Patient Reported Outcomes Research, Department of Applied Health Sciences, University of Birmingham, Edgbaston, Birmingham, UK.
Rationale & Objective: Valid measures of side effects are important to inform clinical use of calcineurin inhibitors (CNIs). This study sought to develop and establish the content validity of a PRO measure to capture side effects among kidney transplant recipients taking CNIs.
Study Design: Qualitative interviews for concept elicitation and cognitive debriefing.
Clin Nutr ESPEN
January 2025
Department of Gastroenterology and Hepatology, Intestinal Failure Unit, Radboud University Medical Centre Nijmegen, Geert Grooteplein 10, 6500 HB, Nijmegen, The Netherlands. Electronic address:
Background And Aims: Measurement of the urine sodium concentration (USC) is a simple procedure that in many patients adequately indicates their hydration status. This is of particular importance in patients suffering from short bowel syndrome (SBS), who may very rapidly dehydrate and are at risk for permanently compromising their kidney function. A point of care test (POCT) that allows reliable measurement of USC would enable these patients to effectively evaluate their sodium- and water balance in the at home setting, thereby avoiding hospital visits and delayed test results.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!
© LitMetric 2025. All rights reserved.