Purpose: Technetium-99m (99mTc) dimercaptosuccinic acid (DMSA) single-photon emission computed tomography (SPECT) has been used to diagnose renal scarring. The Japanese Society of Nuclear Medicine recently revised the 'Consensus Guidelines for Pediatric Nuclear Medicine Examination.' In this study, we compared simulation data with actual data obtained using a pediatric phantom for 99mTc-DMSA examinations and evaluated the usefulness of simulations in determining the optimal acquisition conditions for SPECT images.
Methods: A SPECT quality assurance (QA) phantom study produced images with a renal-to-background 99mTc ratio of 283:1 kBq/ml. The projection data for the simulation were simulated using the simulation of imaging nuclear detectors. To compare the actual measurements and simulations, recovery factors were used for the SPECT QA phantom for image quality assessment. Defect contrast and visual evaluation using Scheffe's method of pairwise comparison were used for the pediatric kidney phantom.
Results: The optimal imaging settings using a kidney phantom required an acquisition time of more than 8 min. The maximum difference in the recovery coefficient between the simulation and actual measurement using the SPECT QA phantom was 6%.
Conclusion: We showed that an acquisition time of more than 8 min was necessary for DMSA-SPECT. In addition, phantom simulations were approximately equivalent to the actual measurement data and the adaptability of simulations was confirmed.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MNM.0000000000001597 | DOI Listing |
Cancer Imaging
January 2025
Melbourne Theranostic Innovation Centre, Level 8, 14-20 Blackwood St, North Melbourne, VIC, 3051, Australia.
True total-body and extended axial field-of-view (AFOV) PET/CT with 1m or more of body coverage are now commercially available and dramatically increase system sensitivity over conventional AFOV PET/CT. The Siemens Biograph Vision Quadra (Quadra), with an AFOV of 106cm, potentially allows use of significantly lower administered radiopharmaceuticals as well as reduced scan times. The aim of this study was to optimise acquisition protocols for routine clinical imaging with FDG on the Quadra the prioritisation of reduced activity given physical infrastructure constraints in our facility.
View Article and Find Full Text PDFBMC Med Educ
January 2025
Department of Forensic Medicine, Karnali Academy of Health Sciences, Chandannath, Jumla, 21200, Nepal.
Following the establishment of Nepal's first medical college in 1972, forensic medicine was introduced in 1978. To date, 25 medical colleges in the country have included forensic medicine as a compulsory subject in the undergraduate medical curriculum. Although this subject has been introduced into the medical curriculum, the outcome is unsatisfactory, as reflected by the poor medico-legal reports prepared by newly graduated medical students.
View Article and Find Full Text PDFInsights Imaging
January 2025
Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.
Objectives: This article aims to evaluate the use and effects of an artificial intelligence system supporting a critical diagnostic task during radiology resident training, addressing a research gap in this field.
Materials And Methods: We involved eight residents evaluating 150 CXRs in three scenarios: no AI, on-demand AI, and integrated-AI. The considered task was the assessment of a multi-regional severity score of lung compromise in patients affected by COVID-19.
Int Ophthalmol
January 2025
Ophthalmology Department, Faculty of Medicine - Suez Canal University Hospitals, 4.5 Kilo - Ring Road, Ismailia, 41522, Egypt.
Purpose: To assess the effect of anterior chamber depth on corneal endothelium using specular microscopy following uneventful phacoemulsification among cataract patients with different axial lengths.
Methods: The study was conducted in a quasi-experimental design including 300 eyes of 300 patients with grade three age-related nuclear cataract distributed equally based on their axial length into three equal groups. All eyes had grade three nuclear cataract.
Sci Rep
January 2025
Department of Neurosurgery and Neurooncology, First Faculty of Medicine, Charles University and Military University Hospital, U Vojenske nemocnice 1200, Prague, 169 02, Czech Republic.
The histological grade is crucial for therapeutic management, and its reliable preoperative detection can significantly influence treatment approach. Lacking established risk factors, this study identifies preoperative predictors of high-grade skull base meningiomas and discusses the implications of non-invasive detection. A multicentric study was conducted on 552 patients with skull base meningiomas who underwent primary surgical resection between 2014 and 2019.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!