Back Plate Marking of a Mechanical Chest Compression Device to Reduce the Duration of Chest Compression Interruptions.

Open Access Emerg Med

Department of Emergency Medicine, Songklanagarind Hospital, Faculty of Medicine, Prince of Songkhla University, Hat Yai, Songkhla, Thailand.

Published: August 2022

Objective: To compare the effectiveness of applying the back plate marking method vs the standard method, to a mechanical chest compression device, in regards to reducing the duration of chest compression interruptions during a simulated cardiac arrest.

Methods: An experimental study, one group pretest posttest design, conducted in a university-based hospital from November 2020 to October 2021. The study recruited 20 participants including emergency medical residents and paramedics. The participants were randomized into three-person teams and applied the device in both standard and back plate marking methods in sequential order. Teams were required to use a mechanical chest compression device in a manikin-based OHCA simulation to assess performance.

Results: The median time pause for the deployment of the upper part of the device was significantly reduced (16 vs 21s, P < 0.01) in the back plate marking method, as was the total pause for device deployment (31.5 vs 38.75s, P = 0.03) and the proportion of total hands-off time attributable to device application interruption (43.08% vs 49.18%, P = 0.02). There was no difference between groups in the duration of all compression interruptions (70.5 vs 82.75s, P = 0.20) and compression fractions (77.85 vs 76.91%, P = 0.19).

Conclusion: The back plate marking method was a significantly reduced time of the deployment of the upper part of the device and in regards to the overall pause for device deployment, but there was no difference in CPR quality between the two methods.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9356708PMC
http://dx.doi.org/10.2147/OAEM.S368510DOI Listing

Publication Analysis

Top Keywords

plate marking
20
chest compression
20
mechanical chest
12
compression device
12
compression interruptions
12
marking method
12
device
9
duration chest
8
deployment upper
8
upper device
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!