Metal artifact reduction in cervix brachytherapy with titanium applicators using dual-energy CT through virtual monoenergetic images and an iterative algorithm: A phantom study.

Brachytherapy

Ionizing and Non-Ionizing Radiation Protection Research Center (INIRPRC), School of Paramedical Sciences, Shiraz University of Medical Sciences, Shiraz, Iran; Physics Unit, Department of Radio-oncology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran. Electronic address:

Published: December 2022

Purpose: To evaluate an iterative metal-artifact reduction (iMAR) algorithm, dual-energy CT (DECT) through virtual monoenergetic images (VMI), and a combination of iMAR and DECT for reducing metal artifact severity (AS) induced by Fletcher titanium applicators used in cervix brachytherapy, the efficacy of which are hitherto unreported.

Methods And Materials: 120 kV single-energy CT (SECT) (Siemens) of BEBIG tandem applicators, varying in shape (straight or curved) and diameter (3.5 mm or 5 mm) in a custom-made water-filled phantom, and their DECT images obtained from extrapolation of 80 kVp and 140 kVp, were reconstructed using four methods: DECT through VMI±iMAR, and SECT±iMAR. The DECT images were reconstructed monoenergetically at 70, 150, and 190 keV. AS was evaluated using measured values and statistical analysis.

Results: iMAR, DECT, and combined DECT and iMAR reduced AS (p < 0.05). DECT had a lower AS than SECT, even without iMAR (p < 0.025). SECT+iMAR was more effective than DECT-iMAR with VMI at 70 and 190 keV (p < 0.05), whereas showing no statistically significant difference at 150 keV. With DECT and iMAR combined, AS was reduced more effectively compared to the SECT+iMAR or DECT alone. It also reduced the mean interobserver uncertainty by 0.2 mm.

Conclusions: These findings indicate that iMAR reduces the AS caused by Fletcher titanium applicators for both SECT and DECT, a combination of iMAR and DECT is superior to either strategy alone, and at low energies, DECT+iMAR also produces similar artifact reduction. These practical strategies promise more accurate source-position and structure definitions in CT-based gynecological brachytherapy treatment planning.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.brachy.2022.07.003DOI Listing

Publication Analysis

Top Keywords

titanium applicators
12
dect
12
imar dect
12
metal artifact
8
artifact reduction
8
cervix brachytherapy
8
virtual monoenergetic
8
monoenergetic images
8
imar
8
combination imar
8

Similar Publications

Article Synopsis
  • - The study evaluated the effectiveness of the AcurosBV dose calculation algorithm for high-dose-rate (HDR) brachytherapy in patients with locally advanced cervical cancer, comparing it against the standard TG43 method using tandem and ovoid applicators.
  • - A cohort of 32 patients received HDR brachytherapy with a prescribed dose, and results showed that AcurosBV provided lower dosimetric values for clinical target volumes (CTV) and organs at risk (OARs) compared to TG43.
  • - The findings indicated an average reduction of approximately 4.3% for CTV dosimetry and varying reductions for OARs, suggesting that AcurosBV is a preferable option for treatments with high heterogeneity in bra
View Article and Find Full Text PDF

Purpose: The design of the angulated screw channel in implant restorations allows the possibility to correct angulation discrepancies, especially in the anterior maxilla. However, the effects of varied screw channel angulations on fracture resistances and fracture patterns of the implant restorations are still uncertain, and thus the aim of this study.

Materials And Methods: Angulated screw channel monolithic zirconia crowns (Nobel Biocare) with three different angulation groups-straight (ASC1), 15° (ASC15), and 25° (ASC25)-were digitally designed from a left central incisor prototype scan.

View Article and Find Full Text PDF

Purpose: Suboptimal ultrasound conspicuity of the brachytherapy applicator can lead to inaccurate image reconstructions of the applicator resulting in decreased tumor control or increased normal tissue dose. This feasibility study aims to improve ultrasound conspicuity of high-dose rate (HDR) brachytherapy needles by modifying the surface of the needles to produce a color Doppler twinkling signature.

Materials And Methods: Surface modifications of standard 17-gauge titanium HDR brachytherapy needles included laser-scribing, application of polymethyl methacrylate (PMMA), and coating with a commercially available echogenic coating.

View Article and Find Full Text PDF

Injectable and photocurable biomaterials are receiving a lot of attention due to their ease of application syringe or dedicated applicator and ability to be used in laparoscopic and robotic minimally invasive procedures. The aim of this work was to synthesize photocurable ester-urethane macromonomers using a heterometallic magnesium-titanium catalyst, magnesium-titanium(iv) butoxide for elastomeric polymer networks. The progress of the two-step synthesis of macromonomers was monitored using infrared spectroscopy.

View Article and Find Full Text PDF

Low-energy X-ray sources that operate in the kilovoltage energy range have been shown to induce more cellular damage when compared to their megavoltage counterparts. However, low-energy X-ray sources are more susceptible to the effects of filtration on the beam spectrum. This work sought to characterize the biological effects of the Xoft Axxent® source, a low-energy therapeutic X-ray source, both with and without the titanium vaginal applicator in place.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!