Objective: The aim of this work was to study the accuracy of the intraoral scanner for detection of tooth wear in natural teeth by using micro-computed tomography (micro-CT) as a gold standard.

Materials And Methods: Twenty premolars were prepared, fixed in acrylic blocks, and scanned with an intraoral scanner (iTero Element® 2) and micro-CT for baseline reference images before artificial tooth wear induction. The samples were then scrubbed with abrasive sandpaper 20 times and scanned with the intraoral scanner. They were then superimposed with the reference images utilising the "TimeLapse" feature of the scanner until the abraded area appeared yellow, indicating tooth surface loss in the 50-200 μm range. The same samples were then rescanned by micro-CT to measure the actual tooth surface loss. This procedure was repeated for the subsequent experimental tooth surface loss of 200-400 μm range (orange areas) and 400-750 μm range (red areas). The collected data were analysed for sensitivity, positive predictive value (PPV), and accuracy. Level of statistical significance was set at .05.

Results: In the detection of experimental tooth surface loss, the specificity, PPV, and accuracy of the intraoral scanner were 98%, 98%, and 97%, respectively.

Conclusions: The iTero® intraoral scanner can be recommended to be a suitable screening tool for tooth wear in routine dental practice.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9875261PMC
http://dx.doi.org/10.1016/j.identj.2022.06.004DOI Listing

Publication Analysis

Top Keywords

intraoral scanner
24
tooth wear
16
tooth surface
16
surface loss
16
accuracy intraoral
12
μm range
12
scanner detection
8
tooth
8
detection tooth
8
scanned intraoral
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!