Optimized Camera-Based Patient Positioning in CT: Impact on Radiation Exposure.

Invest Radiol

From the Department of Radiology and Nuclear Medicine, Klinikum Nuernberg, Paracelsus Medical University, Nuernberg.

Published: February 2023

Objective: The aim of this study was to evaluate whether a 3-dimensional (3D) camera can outperform highly trained technicians in precision of patient positioning and whether this transforms into a reduction in patient exposure.

Materials And Methods: In a single-center study, 3118 patients underwent computer tomography (CT) scans of the chest and/or abdomen on a latest generation single-source CT scanner supported with an automated patient positioning system by 3D camera. One thousand five hundred fifty-seven patients were positioned laser-guided by a highly trained radiographer (camera off) and 1561 patients with 3D camera (camera on) guidance. Radiation parameters such as effective dose, organ doses, CT dose index, and dose length product were analyzed and compared. Isocenter accuracy and table height were evaluated between the 2 groups.

Results: Isocenter positioning was significantly improved with the 3D camera ( P < 0.001) as compared with visual laser-guided positioning. Absolute table height differed significantly ( P < 0.001), being higher with camera positioning (165.6 ± 16.2 mm) as compared with laser-guided positioning (170.0 ± 20.4 mm). Radiation exposure decreased using the 3D camera as indicated by dose length product (321.1 ± 266.6 mGy·cm; camera off: 342.0 ± 280.7 mGy·cm; P = 0.033), effective dose (3.3 ± 2.7 mSv; camera off: 3.5 ± 2.9; P = 0.053), and CT dose index (6.4 ± 4.3 mGy; camera off: 6.8 ± 4.6 mGy; P = 0.011). Exposure of radiation-sensitive organs such as colon ( P = 0.015) and red bone marrow ( P = 0.049) were also lower using the camera.

Conclusions: The introduction of a 3D camera improves patient positioning in the isocenter of the scanner, which results in a lower and also better balanced dose reduction for the patients.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0000000000000904DOI Listing

Publication Analysis

Top Keywords

patient positioning
16
camera
12
positioning
8
radiation exposure
8
highly trained
8
effective dose
8
dose length
8
length product
8
table height
8
laser-guided positioning
8

Similar Publications

Background: Despite progress made towards SDG 3, sub-Saharan Africa lags behind the rest of the world, accounting for over 50% of global neonatal deaths. The increased number of hospital births in the region has not reciprocated the reduction in neonatal mortality rates. Sick newborns face uncertain journeys from peripheral facilities to specialized centres arriving in suboptimal conditions, which impacts their outcomes, due partly to the scarcity of dedicated neonatal transport services.

View Article and Find Full Text PDF

Background: Total hip arthroplasty is the preferred treatment for advanced hip osteoarthritis, yet complications like hip dislocation (0.2 %-10 %) persist due to factors such as implant design, positioning, surgical technique, and patient-specific conditions. Impingement between prosthetic components or the acetabulum and proximal femur is a primary cause of instability.

View Article and Find Full Text PDF

Purpose: Achieving precise postoperative alignment is critical for the long-term success of total knee arthroplasty (TKA). Long-leg standing radiograph (LLR) at 6 weeks post-op is the gold standard for assessing alignment, but its reliance on weight-bearing and positioning makes it less practical in the early postoperative period. Supine computed tomography scanogram (CTS) offers a potential alternative.

View Article and Find Full Text PDF

Background: The number of patients with neurological disorders and severe disability is increasing globally. These patients often need help with positioning and the amount of support varies with their level of impairment. High rates of work-related musculoskeletal disorders are observed among healthcare professionals (HCP) with patient contact due to injuries during manual handling.

View Article and Find Full Text PDF

Full-arch zirconia restorations on implants have gained popularity due to zirconia's strength and aesthetics, yet they are still associated with challenges like structural fractures, peri-implant complications, and design misfits. Advances in CAD/CAM and digital workflows offer potential improvements, but a technique that consistently addresses these issues in fixed, full-arch, implant-supported prostheses is needed. This novel technique integrates a facially and prosthetically driven treatment approach, which is divided into three phases: data acquisition, restoration design, and manufacturing/delivery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!