Background: Irreversible electroporation (IRE) ablation is generally performed with multielectrode catheters. Electrode-tissue contact is an important predictor for the success of pulmonary vein (PV) isolation; however, contact force is difficult to measure with multielectrode ablation catheters. In a preclinical study, we assessed the feasibility of a multielectrode impedance system (MEIS) as a predictor of long-term success of PV isolation. In addition, we present the first-in-human clinical experience with MEIS.
Methods: In 10 pigs, one PV was ablated based on impedance (MEIS group), and the other PV was solely based on local electrogram information (EP group). IRE ablations were performed at 200 J. After 3 months, recurrence of conduction was assessed. Subsequently, in 30 patients undergoing PV isolation with IRE, MEIS was evaluated and MEIS contact values were compared to local electrograms.
Results: In the porcine study, 43 IRE applications were delivered in 19 PVs. Acutely, no reconnections were observed in either group. After 3 months, 0 versus 3 (=0.21) PVs showed conduction recurrence in the MEIS and EP groups, respectively. Results from the clinical study showed a significant linear relation was found between mean MEIS value and bipolar dV/dt (r=0.49, <0.001), with a slope of 20.6 mV/s per Ohm.
Conclusions: Data from the animal study suggest that MEIS values predict effective IRE applications. For the long-term success of electrical PV isolation with circular IRE applications, no significant difference in efficacy was found between ablation based on the measurement of electrode interface impedance and ablation using the classical EP approach for determining electrode-tissue contact. Experiences of the first clinical use of MEIS were promising and serve as an important basis for future research.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9384826 | PMC |
http://dx.doi.org/10.1161/CIRCEP.121.010835 | DOI Listing |
J Interv Card Electrophysiol
January 2025
Department of Cardiology, Thorax Center, Cardiovascular Institute, Erasmus MC, Rotterdam, the Netherlands.
Introduction: A hybrid approach with very high-power short-duration (vHPSD) posteriorly and ablation-index guided HPSD (50 W) anteriorly seems to be an optimal balance between efficiency and effectiveness for point-by-point pulmonary vein isolation (PVI). The aim of the current study is to compare vHPSD/HPSD ablation to cryoballoon ablation (CBA) in patients with symptomatic atrial fibrillation (AF).
Methods And Results: In this retrospective single-center study, we identified 110 consecutive patients who underwent their first PVI with either vHPSD/HPSD (n = 54) or CBA (n = 56).
J Clin Med
January 2025
Department of Pathophysiology, George Emil Palade University of Medicine, Pharmacy, Science, and Technology of Targu Mures, 540142 Targu Mures, Romania.
Venous thromboembolism (VTE), encompassing deep vein thrombosis and pulmonary embolism, is a significant burden on health and economic systems worldwide. Improved VTE management calls for the integration of biomarkers into diagnostic algorithms and scoring systems for risk assessment, possible complications, and mortality. This literature review discusses novel biomarkers with potential diagnostic and prognostic value in personalized VTE management.
View Article and Find Full Text PDFJ Clin Med
December 2024
Department of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, 20072 Pieve Emanuele, Italy.
Pulmonary vein isolation (PVI) represents the cornerstone of paroxysmal (PAF) and persistent atrial fibrillation (PsAF) ablation. Impedance values provide insights on tissue conductive properties. Consecutive patients undergoing PAF and PsAF ablation were prospectively enrolled.
View Article and Find Full Text PDFJ Clin Med
December 2024
Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum Der Charité, 13353 Berlin, Germany.
: Atrial fibrillation (AF) is a common cardiac arrhythmia associated with left atrial dysfunction. The impact of pulmonary vein isolation (PVI) using pulsed field ablation (PFA) on left atrial function has not been previously quantified. This study aims to evaluate the effects of PVI using PFA on left atrial function in patients with AF.
View Article and Find Full Text PDFDiagnostics (Basel)
January 2025
Department of Radiology, Azienda Ospedaliero Universitaria, University of Cagliari, 09124 Cagliari, Italy.
Over the past decade, several trials and observational studies have validated the use of minimally invasive cardiac interventions as viable treatment options for various cardiac diseases. Transcatheter techniques for severe aortic valve stenosis have rapidly emerged as alternatives to surgical aortic valve replacement in certain patient populations. Additionally, non-surgical treatment options have expanded for conditions affecting other cardiac valves, such as the mitral valve.