Objective: This study aimed to evaluate the safety and efficacy of the fixation of transforaminal sacral fractures using TiRobot-assisted transiliac-transsacral (TITS) screws under multimodal neuroelectrophysiological monitoring (MNM).
Methods: From January 2019 to May 2021, 22 patients (17 male and 5 female patients) with transforaminal sacral fractures who were treated with closed reduction and placement of TiRobot-assisted TITS screws under MNM were retrospectively evaluated. The average age of the patients was 43.32 ± 11.40 years (range: 19-63). The patients received MNM, including somatosensory-evoked potentials (SEPs), motor-evoked potentials (MEPs), and electromyographic monitoring (EMG), prior to surgery, during closed reduction and the placement of the guidewire and TITS screw, and at the end of surgery. The operation was adjusted according to the MNM results.
Results: Overall, 22 TITS screws were inserted in 22 patients, including 5 TITS screws in the S1 body and 17 TITS screws in the S2 body. The average time needed for screw placement was 27.95 ± 6.84 mins, and the average frequency of X-ray fluoroscopy exposures was 31.00 ± 5.56 for each patient. Anterior ring fixation was performed in 4 patients using an external fixator, in 5 patients using cannulated screws, and in 13 patients using reconstruction plates. The mean follow-up time was 14.46 ± 2.46 months (12-20 months). Tornetta and Matta radiographic outcomes were excellent in 10 patients, good in 9 patients, fair in 2 patients, and poor in 1 patient. The proportion of excellent and good ratings was 86.36%. At the final follow-up, the average Majeed score was 82.18 ± 14.52, with clinical outcomes that were excellent in 9 patients, good in 9 patients, fair in 1 patient, and poor in 3 patients. The proportion of excellent and good ratings was 82.82%. Preoperatively, the amplitude of the SEP on the injured side was lower than that on the contralateral side before reduction in 9 patients (>50%). In this study, no screw was mistakenly inserted into the sacral canal, and no surgical site infection occurred.
Conclusion: MNM combined with TiRobot assistance can safely implant TITS screws and can effectively identify the neurological function of patients under anesthesia and reduce iatrogenic nerve injury.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9338859 | PMC |
http://dx.doi.org/10.1155/2022/3383665 | DOI Listing |
Orthop Surg
December 2024
Department of Orthopaedics, The Fourth Medical Center, Chinese PLA General Hospital, Beijing, China.
Objective: There are many advantages to stabilize the posterior pelvic ring injuries with a transiliac-transsacral (TITS) screw percutaneously. To identify the correct entry point and insert a guidewire accurately for a TITS screw, we propose a method of specifying the optimal entry point, and introduce a technique of enabling freehand placement of a guidewire with fluoroscopic guidance.
Methods: In this retrospective study, 116 patients who underwent pelvic CT scans and pelvic lateral radiographs at our institution from January 2020 to April 2022 were enrolled.
J Orthop Traumatol
September 2023
Division of Orthopedic Traumatology, Department of Orthopedic Surgery, Chang Gung Memorial Hospital, No. 5, Fu-Hsing Street, Kweishan 333, Taoyuan, Taiwan.
Background: Percutaneous iliosacral (IS) screw fixation and trans-iliac trans-sacral (TITS) screw fixation are clinically effective treatments of posterior pelvic sacroiliac fractures. In order to accurately assess the sacrum position relative to the pelvis, pelvic incidence (PI) is a commonly utilized radiographic parameter in sagittal view. This study aimed to investigate and compare the surgical outcomes and radiographic parameters of IS or TITS screw fixations for the treatment of posterior sacroiliac complex fractures with different PI values.
View Article and Find Full Text PDFEur J Trauma Emerg Surg
December 2023
Department of Trauma-, Hand-, and Reconstructive Surgery, Ulm University, Albert-Einstein-Allee 23, 89081, Ulm, Germany.
Introduction: Pelvic fractures were often associated with high-energy trauma in young patients, but data show a significant increase in osteoporotic pelvic fractures in old age due to the progressive demographic change. There is an ongoing discussion about the best fixation techniques, which are ranging from lumbopelvic fixation to sacral bars or long transiliac-transsacral (TITS) screws. This study analyzes TITS screw osteosynthesis and sacroiliac screw osteosynthesis (SI), according to biomechanical criteria of fracture stability in osteoporotic human pelvic cadavers ex vivo.
View Article and Find Full Text PDFArch Orthop Trauma Surg
October 2023
Department of Orthopedics, McGovern Medical School at UTHealth, 6400 Fannin St, Suite 1700, Houston, TX, 77030, USA.
Introduction: The purpose of this study is to (1) describe a pre-operative planning technique using non-reformatted CT images for insertion of multiple transiliac-transsacral (TI-TS) screws at a single sacral level, (2) define the parameters of a sacral osseous fixation pathway (OFP) that will allow for insertion of two TI-TS screws at a single level, and (3) identify the incidence of sacral OFPs large enough for dual-screw insertion in a representative patient population.
Methods: Retrospective review at a level-1 academic trauma center of a cohort of patients with unstable pelvic injuries treated with two TI-TS screws in the same sacral OFP, and a control cohort of patients without pelvic injuries who had CT scans for other reasons.
Results: Thirty-nine patients had two TI-TS screws at S1.
Biomed Res Int
August 2022
Department of Orthopedics and Traumatology, Xi'an Honghui Hospital, Xi'an JiaoTong University Health Science Center, Xi'an, China.
Objective: This study aimed to evaluate the safety and efficacy of the fixation of transforaminal sacral fractures using TiRobot-assisted transiliac-transsacral (TITS) screws under multimodal neuroelectrophysiological monitoring (MNM).
Methods: From January 2019 to May 2021, 22 patients (17 male and 5 female patients) with transforaminal sacral fractures who were treated with closed reduction and placement of TiRobot-assisted TITS screws under MNM were retrospectively evaluated. The average age of the patients was 43.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!