Background: In atrial fibrillation (AF) patients, catheter ablation of pulmonary veins (PVI) is the most effective therapeutic option to maintain sinus rhythm. To improve successful PVI, contact force-sensing (CF) catheters became routinely available. Previous studies did not clearly show superior clinical efficacy in comparison with non-CF catheters.
Methods: We investigated consecutive patients, who underwent index PVI for AF at our hospital between 2012 and 2018. Three hundred and fifty-four patients were ablated without CF. After availability of CF catheters in 2016, 317 patients were ablated using CF. In case of crossover between the groups, follow-up was censored. The primary endpoint was any documented atrial tachycardia (AT) or atrial fibrillation > 30 s after a 3-month blanking period. Secondary endpoints were procedural characteristics and periprocedural complications.
Results: There was no significant difference between the groups at baseline except hyperlipidemia. After 365 days of follow-up, 67% of patients in the CF group remained free from AF/AT recurrence compared to 59% in non-CF group (P = 0.038). In multivariable Cox regression analysis, non-CF ablation was an independent risk factor for AF recurrence besides age and persistent AF. Total fluoroscopy time (15 ± 7.6 vs. 28 ± 15.9 min) and total procedure time (114 ± 29.6 vs. 136 ± 38.5 min) were significantly lower for CF-guided PVI (P < 0.001). Complication rates did not differ between groups (P = 0.661).
Conclusions: In our study, the AT/AF recurrence rate and pulmonary vein reconnection rate is lower after CF PVI with a similar complication rate but lower total procedure time and total fluoroscopy time compared to non-CF PVI.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9726666 | PMC |
http://dx.doi.org/10.1007/s10840-022-01316-8 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!