Objective: The purpose of this study was to provide an updated estimate of the prevalences of different types of human papillomavirus (HPV) in females in Chaoshan District and to establish an internal quality control (IQC) method for excluding false-positive results in HPV detection by using the Levey-Jennings control chart.

Method: HPV types were detected in 23,762 cervical samples by using PCR membrane hybridization. The means and standard deviations (SDs) of the positive rates were calculated, the Levey-Jennings chart was plotted, and the rules for "out of control" and "warning" were established. A set of standardized IQC for HPV DNA tests was developed based on the values and Levey-Jennings charts.

Result: In 466 batches, the positive rate exceeded the 1 + 2SD rule 24 times, but there was no consecutive exceedance, which was considered "in control". When the positive rate exceeded the 1 + 3SD rule 8 times with consecutive exceedance, it was considered "out of control". Further examination revealed that detections showing "out of control" had an undesirable random error, indicating that contamination may occur due to improper operation.

Conclusion: This unique Levey-Jennings control chart is a practical method for eliminating false-positive results in HPV DNA detection and should be widely applicable in molecular diagnostic laboratories.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9331565PMC
http://dx.doi.org/10.1186/s12985-022-01861-8DOI Listing

Publication Analysis

Top Keywords

levey-jennings control
12
"out control"
12
unique levey-jennings
8
control chart
8
internal quality
8
quality control
8
human papillomavirus
8
false-positive hpv
8
hpv dna
8
positive rate
8

Similar Publications

To ensure the quality of their analyses, medical laboratories carry out internal quality control (IQC) on a daily basis. IQC involves control samples with known target values for all parameters used by a laboratory in clinical practice. The use of IQC enables the laboratory to monitor the accuracy and precision of laboratory results.

View Article and Find Full Text PDF

Background And Aim: This study applied Six Sigma metrics to facilitate the quality control (QC) review for hospital glucose meters.

Materials And Methods: QC data from a period of six months on all hospital glucose meters were extracted from the data management system. Sigma values for each meter at two QC levels were calculated and evaluated each month by combining the imprecision, the absolute bias between the meter mean and all-meter mean, and the standards from ISO 15179:2013.

View Article and Find Full Text PDF

Introduction: Quality Control Management (QCM) in clinical laboratories is crucial for ensuring reliable results in analytical measurements, with biological variation being a key factor. The study focuses on assessing the analytical performance of the Reverse Transcription Polymerase Chain Reaction (RT-PCR) system for Human Immunodeficiency Virus (HIV), Hepatitis B (HBV), and Hepatitis C (HCV). Five models proposed between 1999 and 2014 offer different approaches to evaluating analytical quality, with Model 2 based on biological variation and Model 5 considering the current state of the art.

View Article and Find Full Text PDF

The influence of calibration on bias in quality control and patient results for TSH on Vitros XT 7600 analyzer.

Biochem Med (Zagreb)

October 2024

Hamilton Regional Laboratory Medicine Program (HRLMP), Hamilton, Ontario, Canada.

Introduction: Thyroid-stimulating hormone (TSH) is a glycoprotein secreted by the anterior pituitary gland and is regulated by negative feedback from the serum free thyroid hormones. In this study we aimed to quantitate the relative bias caused by calibration drifting as seen in our TSH Levey-Jennings quality control (QC) charts and assess the magnitude of bias on patients' samples.

Materials And Methods: In the period from October 2021 to August 2022 we looked at the QC results of ten 28-days' calibration time intervals and calculated relative bias compared to the mean.

View Article and Find Full Text PDF

QC Constellation: a cutting-edge solution for risk and patient-based quality control in clinical laboratories.

Clin Chem Lab Med

October 2024

General Directorate of Health Services, Rare Diseases Department, Turkish Ministry of Health, Ankara, Türkiye.

Objectives: Clinical laboratories face limitations in implementing advanced quality control (QC) methods with existing systems. This study aimed to develop a web-based application to addresses this gap, and improve QC practices.

Methods: QC Constellation, a web application built using Python 3.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!