Purpose: To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.

Patients And Methods: A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18-50 years and a corrected Snellen's distance visual acuity (CDVA) of 0.8 decimal or better. Angle Kappa is assessed directly using Orbscan® III software version 1.8.165.1. (Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.21r.65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.2.1.60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.

Results: Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (- 0.18° ± 1.8), and it was statistically insignificant (p value = 0.1294). Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (p value = 0.0004 and < 0.0001 consecutively). Bland Altman analysis showed a 95% confidence interval between Orbscan III and Pentacam of - 3.76 to 3.4 and between Orbscan III and Sirius of - 3.79 to 2.26.

Conclusion: Pentacam parameters can be used as a reliable method to calculate angle kappa indirectly, without usage of any additional measurements from other machine. Sirius device parameters could also be used, but with less accurate results. A simple modification to those devices' software to calculate it, and incorporate it in the printout is possible, and highly recommended.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9971106PMC
http://dx.doi.org/10.1007/s10792-022-02433-zDOI Listing

Publication Analysis

Top Keywords

angle kappa
20
pentacam sirius
8
angle
5
kappa agreement
4
agreement scheimpflug
4
scheimpflug tomography
4
tomography combined
4
combined placido
4
placido scheimpflug
4
scheimpflug combined
4

Similar Publications

Purpose: To assess the relationship between angle kappa (apparent chord mu) and ocular parameters in cataract patients.

Methods: In this cross-sectional study, the data on apparent chord mu, age, axial length, anterior chamber depth, anterior and posterior mean keratometry, mean total keratometry, white-to-white, central corneal thickness and lens thickness were collected for consecutive cataract patients. Correlation (Pearson) between chord mu and the other ocular parameters was calculated.

View Article and Find Full Text PDF

Purpose: To propose a new sign of patellar maltracking in recurrent patellar dislocation (RPD) and compare the differences in lower limb rotational and bony structural abnormalities among the different signs.

Patients And Methods: A retrospective study included 279 patients (mean age: 22 years; female: 81%) who underwent primary surgery for RPD over the past 4 years was performed. The patients were grouped based on the characteristics of patellar tracking: low-, moderate- and high-grade J-sign.

View Article and Find Full Text PDF

: Idiopathic scoliosis (IS) is a common spinal deformity affecting 0.5% to 5.2% of children worldwide, with a higher reported range in Spain (0.

View Article and Find Full Text PDF

Digital Health Technologies for Optimising Treatment and Rehabilitation Following Surgery: Device-Based Measurement of Sling Posture and Adherence.

Sensors (Basel)

December 2024

Assessment of Movement Behaviours (AMBer), Leicester Lifestyle and Health Research Group, Diabetes Research Centre, University of Leicester, Leicester LE5 4PW, UK.

Background: Following shoulder surgery, controlled and protected mobilisation for an appropriate duration is crucial for appropriate recovery. However, methods for objective assessment of sling wear and use in everyday living are currently lacking. In this pilot study, we aim to determine if a sling-embedded triaxial accelerometer and/or wrist-worn sensor can be used to quantify arm posture during sling wear and adherence to sling wear.

View Article and Find Full Text PDF

Purpose: Correctly classifying uterine fibroids is essential for treatment planning. The objective of this study was to assess the accuracy and reliability of the FIGO classification system in categorizing uterine fibroids via organ-axial T2WI and to further investigate the factors associated with uterine compression.

Methods: A total of 130 patients with ultrasound-confirmed fibroids were prospectively enrolled between March 2023 and May 2024.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!