Background: The aim of this study was to redefine the icterus index cutoff for the Roche Jaffé creatinine method using both conjugated and unconjugated bilirubin on 3 Roche cobas modules (c311, c501, and c701/c702) at laboratories across our hospital network.

Methods: Interference was evaluated by adding conjugated bilirubin (as bilirubin conjugate, ditaurate) and unconjugated bilirubin to pooled remnant plasma. The effects of conjugated and unconjugated bilirubin were tested separately to assess the contribution of each species. The magnitude of interference was calculated as both absolute and percentage error with total allowable error limits set at 0.1 mg/dL (8 µmol/L or 8%).

Results: Analysis of interference data across the 3 Roche modules did not show bias exceeding our total allowable error limits for plasma creatinine up to a conjugated bilirubin icterus index of 16.2 (approximately 16.2 mg/dL or 277 μmol/L) or an unconjugated bilirubin icterus index of 18.5 (approximately 18.5 mg/dL or 316 μmol/L), the highest concentrations tested.

Conclusions: Our results demonstrate that the Roche Jaffé method exhibits acceptable performance in the presence of icterus at icterus indexes above the manufacturer's current recommendations of 5 (approximately 5 mg/dL or 86 μmol/L) and 10 (approximately 10 mg/dL or 171 μmol/L) for conjugated and unconjugated bilirubin, respectively. We have updated the icterus index in our hospital system to 16 for conjugated bilirubin and 18 for unconjugated bilirubin.

Download full-text PDF

Source
http://dx.doi.org/10.1093/jalm/jfac045DOI Listing

Publication Analysis

Top Keywords

unconjugated bilirubin
24
conjugated unconjugated
12
conjugated bilirubin
12
bilirubin
10
icterus cutoff
8
jaffé creatinine
8
creatinine method
8
roche jaffé
8
total allowable
8
allowable error
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!