Background: Novel smartwatch-based cuffless blood pressure (BP) measuring devices are coming to market and receive FDA and CE labels. These devices are often insufficiently validated for clinical use. This study aims to investigate a recently CE-cleared smartwatch using cuffless BP measurement in a population with normotensive and hypertensive individuals scheduled for 24-h BP measurement.
Methods: Patients that were scheduled for 24-h ambulatory blood pressure monitoring (ABPM) were recruited and received an additional Samsung Galaxy Watch Active 2 smartwatch for simultaneous BP measurement on their opposite arm. After calibration, patients were asked to measure as much as possible in a 24-h period. Manual activation of the smartwatch is necessary to measure the BP. Accuracy was calculated using sensitivity, specificity, positive and negative predictive values and ROC curves. Bland-Altman method and Taffé methods were used for bias and precision assessment. BP variability was calculated using average real variability, standard deviation and coefficient of variation.
Results: Forty patients were included. Bland-Altman and Taffé methods demonstrated a proportional bias, in which low systolic BPs are overestimated, and high BPs are underestimated. Diastolic BPs were all overestimated, with increasing bias toward lower BPs. Sensitivity and specificity for detecting systolic and/or diastolic hypertension were 83 and 41%, respectively. ROC curves demonstrate an area under the curve (AUC) of 0.78 for systolic hypertension and of 0.93 for diastolic hypertension. BP variability was systematically higher in the ABPM measurements compared to the smartwatch measurements.
Conclusion: This study demonstrates that the BP measurements by the Samsung Galaxy Watch Active 2 show a systematic bias toward a calibration point, overestimating low BPs and underestimating high BPs, when investigated in both normotensive and hypertensive patients. Standards for traditional non-invasive sphygmomanometers are not met, but these standards are not fully applicable to cuffless devices, emphasizing the urgent need for new standards for cuffless devices. The smartwatch-based BP measurement is not yet ready for clinical usage. Future studies are needed to further validate wearable devices, and also to demonstrate new possibilities of non-invasive, high-frequency BP monitoring.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9309348 | PMC |
http://dx.doi.org/10.3389/fcvm.2022.958212 | DOI Listing |
Diabetes Ther
January 2025
Departamento de Endocrinología y Metabolismo, Unidad de Investigación en Enfermedades Metabolicas, Instituto Nacional de Ciencias Médicas y Nutrición, Salvador Zubirán, Mexico City, Mexico.
Introduction: Young adulthood is well documented as being a particularly challenging area of type 1 diabetes (T1D) healthcare. Many young adults with T1D (YAT1D) are distracted from effective disease self-management; T1D healthcare service engagement can be problematic and inconsistent, and high rates of unplanned healthcare contacts prevail. Video conferencing use can facilitate services to be flexible and responsive.
View Article and Find Full Text PDFJ Cardiothorac Surg
January 2025
Department of Cardiology, Shanghai East Hospital, School of Medicine, Tongji University, Shanghai, China.
Background: Continuous Positive Airway Pressure (CPAP) treatment brings more benefits than risks to most coronary heart disease (CHD) patients with obstructive sleep apnea (OSA). However, the pathophysiological mechanism by which CPAP treatment improves the prognosis of patients with CHD and OSA remains unclear. The purpose of this study was to clarify whether CPAP can improve arterial stiffness and inflammatory factor levels in CHD patients with OSA, and to further improve prognosis.
View Article and Find Full Text PDFJ Hum Hypertens
January 2025
Department of Pediatrics and Child Health, University of Ilorin, Ilorin, Nigeria.
Eur J Pharmacol
January 2025
Department of Nephrology, Xin Hua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai, China. Electronic address:
Calcium/calmodulin dependent protein kinase II inhibitor 1 (Camk2n1) is closely associated with a peak logarithm of odds score in quantitative trait loci for systolic blood pressure. Increased Camk2n1 mRNA expression has been specifically observed in the kidneys of hypertension mouse models. However, the precise role of Camk2n1 in the kidney remains unclear.
View Article and Find Full Text PDFJ Affect Disord
January 2025
Division of Neuropsychiatry and Psychosomatics, Department of Neurology, Xuanwu Hospital, Capital Medical University, Beijing Psychosomatic Disease Consultation Center, National Center for Neurological Disorders, National Clinical Research Center for Geriatric Diseases, Beijing, China. Electronic address:
Background: Transcranial alternating current stimulation (tACS) at 77.5 Hz and 15 mA, targeting the forehead and mastoid areas, has proven efficacious in patients with major depressive disorder (MDD) by simultaneously stimulating multiple brain nuclei and regions, many of which are critical for blood pressure regulation. This post hoc analysis aimed to assess the potential blood pressure-lowering effects of 77.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!