Objective: To evaluate effects of loop diameter of a modified Kessler locking-loop (LL) suture on in vitro tensile strength and gapping characteristics of canine flexor tendon repairs.

Sample: 48 cadaveric superficial digital flexor tendons from 24 adult medium- to large-breed dogs.

Procedures: Flexor tendons were randomly assigned to 4 groups (n = 12/group) and repaired with 2-0 polypropylene in a LL pattern with loops measuring 1, 2, 3, or 4 mm in diameter. Biomechanical loads, gap formation between tendon ends, and failure modes were evaluated and compared between groups.

Results: Increasing loop diameter from 1 to 4 mm significantly increased yield (P = .048), peak (P < .001), and failure (P < .001) loads. There were no significant differences in yield, peak, and failure loads between 1- and 2-mm loops. Load to 3-mm gap formation was significantly (P < .001) greater for 4-mm loops, compared with 1-, 2-, and 3-mm loops. Failure mode did not differ significantly among experimental groups, with 46 of 48 (96%) of constructs failing because of suture breakage.

Clinical Relevance: Loop diameter of a LL suture pattern is an important biomechanical variable that influences construct biomechanics of canine tendon suture repairs. Loop diameters > 3 mm are recommended when the size of the tendon allows. Further studies are necessary to determine the in vivo effect of these findings, particularly the effects on tendon blood supply.

Download full-text PDF

Source
http://dx.doi.org/10.2460/ajvr.21.12.0212DOI Listing

Publication Analysis

Top Keywords

loop diameter
16
diameter modified
8
modified kessler
8
kessler locking-loop
8
locking-loop suture
8
suture vitro
8
vitro tensile
8
tensile strength
8
strength gapping
8
gapping characteristics
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!