The biomedical acoustic signal plays an important role in clinical non-invasive diagnosis. In view of the deficiencies in early diagnosis of cardiovascular diseases, acoustic properties of and heart sounds are utilized. In this paper, we propose an integrated concave cilium MEMS heart sound sensor. The concave structure enlarges the area for receiving sound waves to improve the low-frequency sensitivity, and realizes the low-frequency and high-sensitivity characteristics of an MEMS heart sound sensor by adopting a reasonable acoustic package design, reducing the loss of heart sound distortion and faint heart murmurs, and improving the auscultation effect. Finally, experimental results show that the integrated concave ciliated MEMS heart sound sensor's sensitivity reaches -180.6 dB@500 Hz, as compared with the traditional bionic ciliated MEMS heart sound sensor; the sensitivity is 8.9 dB higher. The sensor has a signal-to-noise ratio of 27.05 dB, and has good heart sound detection ability, improving the accuracy of clinical detection methods.