Background: Surgical aortic valve replacement with a rapid deployment valve (RDV) is a relatively recent treatment option. The aim of this study was to compare the hemodynamic performance of balloon-expandable (BE)-RDVs and BE-transcatheter heart valves (THVs) in a high surgical risk and frail-elderly population.

Methods: BE-THVs and BE-RDVs were implanted in 138 and 47 patients, respectively, all older than 75 years and with a Canadian Study of Health and Aging category of 5 or above. Echocardiographic assessment was performed at discharge and six months later.

Results: At discharge, transprosthetic pressure gradients and indexed effective orifice area (iEOA) were similar in both cohorts. At six-month follow-up, BE-RDVs showed lower peak (14.69 vs. 20.86 mmHg; p < 0.001) and mean (7.82 vs. 11.83 mmHg; p < 0.001) gradients, and larger iEOA (1.05 vs. 0.84 cm/m; p < 0.001). Similar findings were also shown considering only small-sized valves. Moderate-to-severe paravalvular leakage was more prevalent in BE-THVs at discharge (14.49 vs. 0.00%; p = 0.032) and, considering exclusively small prostheses, at six months too (57.69 vs. 15.00%; p = 0.014). Nevertheless, BE-THVs determined amelioration in left ventricular ejection fraction (53.79 vs. 60.14%; p < 0.001), pulmonary artery systolic pressure (35.81 vs. 33.15 mmHg; p = 0.042), and tricuspid regurgitation severity (40.58 vs. 19.57%; p = 0.031), from discharge to mid-term follow-up.

Conclusions: BE-RDVs showed better hemodynamic performance, especially when implanted in small annuli. Despite their worse baseline conditions, transcatheter patients still exhibited a greater improvement of their echocardiographic profile at mid-term follow-up.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hjc.2022.07.006DOI Listing

Publication Analysis

Top Keywords

hemodynamic performance
8
performance balloon-expandable
8
rapid deployment
8
balloon-expandable aortic
4
aortic bioprostheses
4
bioprostheses elderly
4
elderly comparison
4
comparison rapid
4
deployment transcatheter
4
transcatheter implantation
4

Similar Publications

Impact of COVID-19 on heart rate variability in post-COVID individuals compared to a control group.

Sci Rep

December 2024

Cardiopulmonary Physiotherapy Laboratory, Physical Therapy Department, Universidade Federal de São Carlos, Rodovia Washington Luiz, São Carlos, SP, 13565-905, Brazil.

This study investigated the impact of mild COVID-19 on HRV in groups stratified by time after infection and to compare to a healthy group of the same age without previous virus infection and without need of hospitalization. This is a cross-sectional study. We divided the sample into four groups: control group (CG) (n = 31), group 1 (G1): ≤6 weeks (n = 34), group 2 (G2): 2-6 months (n = 30), group 3 (G3): 7-12 months (n = 35) after infection.

View Article and Find Full Text PDF

A systematic review of the comparative effects of sound and music interventions for intensive care unit patients' outcomes.

Aust Crit Care

December 2024

Department of Music, Canadian Centre for Ethnomusicology (CCE), Department of Performing Arts, Faculty of Communication and Media Studies, University for Development Studies, Ghana; Department of Music, Faculty of Arts, University of Alberta, 3-98 Fine Arts Building, Edmonton, AB, T6G 2C9, Canada. Electronic address:

Background: Despite syntheses of evidence showing efficacy of music intervention for improving psychological and physiological outcomes in critically ill patients, interventions that include nonmusic sounds have not been addressed in reviews of evidence. It is unclear if nonmusic sounds in the intensive care unit (ICU) can confer benefits similar to those of music.

Objective: The aim of this study was to summarise and contrast available evidence on the effect of music and nonmusic sound interventions for the physiological and psychological outcomes of ICU patients based on the results of randomised controlled trials.

View Article and Find Full Text PDF

Background: Interventricular dyssynchrony derived from the classic non-physiological stimulation (n-PS) of the right ventricle (RV) is a known cause of left ventricular dysfunction (LVDys).

Methods: This was a prospective descriptive single-center study. We analyzed patients who develop LVDys with n-PS, and the results after upgrading to conduction system pacing (CSP).

View Article and Find Full Text PDF

LieRHRV system for remote lie detection using heart rate variability parameters.

Sci Rep

December 2024

Laboratory of Bioelectric and Bioenergetic Systems, Faculty of Biomedical Engineering, Technion-Israel Institute of Technology, Haifa, Israel.

The standard polygraph, or lie detector, is limited by its reliance on average heart rate, subjective examiner interpretation, and the need for direct subject contact. Remote photoplethysmography (rPPG) offers a promising contactless alternative, by using facial videos to extract heart rate variability (HRV). We introduce "LieRHRV," a remote lie detection algorithm based solely on extracted HRV parameters.

View Article and Find Full Text PDF

Background: Predicted heart mass ratio (PHMr) has become the standard donor-recipient size matching method in heart transplantation. While utilization of small PHMr hearts is associated with increased one-year mortality, the underlying mechanisms and time horizon of mortality remain uncertain.

Methods: A single institution analysis of isolated heart transplant recipients (01/2019-7/2022) was performed (N=334).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!