Background: Although transcatheter aortic valve replacement (TAVR) therapy continues to grow, there have been concerns about the occurrence of hypoattenuating leaflet thickening (HALT), which may affect prosthesis function or durability. This study aimed to examine prosthesis frame factors and correlate their extent to the frequency of HALT and clinical outcomes.

Methods: We prospectively examined 565 patients with cardiac computed tomography screening for HALT at 30 days after balloon-expandable SAPIEN3 and self-expanding EVOLUT TAVR. Deformation of the TAVR prostheses, asymmetric prosthesis leaflet expansion, prosthesis sinus volumes, and commissural alignment were analyzed on the postprocedural computed tomography. For descriptive purposes, an index of prosthesis deformation was calculated, with values >1.00 representing relative midsegment underexpansion. A time-to-event model was performed to evaluate the association of HALT with the clinical outcome.

Results: Overall, HALT was present in 21% of SAPIEN3 patients and in 16% of EVOLUT patients at 30 days after TAVR. The occurrence of HALT was directly associated with greater prosthesis frame deformation (<0.001), worse asymmetry of the leaflets (<0.001), and smaller TAVR neosinus volumes (<0.001). These relations were present in both prosthetic types and in all of their size ranges (all <0.05). In multivariable analyses that include clinical variables previously associated with HALT (eg, anticoagulant therapy), variables of TAVR prosthesis deformation remained predictive of HALT. Although HALT was not associated with changes in prosthetic hemodynamics, its presence was associated with the risk of mortality at 1 year, with respect to greater incidences of all-cause mortality (hazard ratio, 2.98 [95% CI, 1.57-5.63]; =0.001), cardiac death (hazard ratio, 4.58 [95% CI, 1.81-11.6]; =0.001), and a composite outcome of all-cause mortality and heart failure hospitalization (hazard ratio, 1.94 [95% CI, 1.14-3.30]; =0.02) with adjustment for age, sex, and comorbidities.

Conclusions: Nonuniform expansion of TAVR prostheses resulting in frame deformation, asymmetric leaflet, and smaller neosinus volume is related to occurrence of HALT in patients who undergo TAVR. These data may have implications for both prosthesis valve design and deployment techniques to improve clinical outcomes for these patients.

Download full-text PDF

Source
http://dx.doi.org/10.1161/CIRCULATIONAHA.121.058339DOI Listing

Publication Analysis

Top Keywords

transcatheter aortic
8
aortic valve
8
hypoattenuating leaflet
8
leaflet thickening
8
prosthesis frame
8
halt clinical
8
computed tomography
8
halt
6
prosthesis
6
deformation
4

Similar Publications

Transcatheter versus Surgical Aortic Valve Replacement in Patients Younger than 65 Years in the US.

J Thorac Cardiovasc Surg

January 2025

Department of Cardiac Surgery, Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Electronic address:

Objective: To characterize trends and outcomes of aortic valve replacement in patients <65 with aortic stenosis between 2013 and 2021.

Methods: This retrospective analysis included 9,557 patients who underwent biological aortic valve replacement in California, New York, and New Jersey from 2013 through 2021. Patients were stratified by approach: transcatheter aortic valve replacement (TAVR) versus surgical aortic valve replacement (SAVR).

View Article and Find Full Text PDF

Background: Evidence regarding the incidence of prosthesis-patient mismatch (PPM) and long-term mortality after transcatheter aortic valve replacement (TAVR) in patients with bicuspid aortic valve stenosis (AS) is scarce.

Objectives: This study sought to assess the incidence and prognostic impact of PPM after TAVR for bicuspid AS compared with that for tricuspid AS.

Methods: In total, 7,393 patients who underwent TAVR were prospectively enrolled in the OCEAN-TAVI (Optimized Catheter Valvular Intervention Transcatheter Aortic Valve Implantation) registry, an ongoing Japanese, multicenter registry.

View Article and Find Full Text PDF

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!