Cardiac troponin I (cTnI) is a standard biomarker for the diagnosis of acute myocardial infarction (AMI). While older, ultra-sensitive cTnI (us-cTnI) assays use the 99th percentile as the reference threshold, newer high-sensitive cTnI (hs-cTnI) assays use the limit of detection or functional sensitivity instead. However, little has been done to systematically compare these two methods. The present study also served as a validation of hs-cTnI in our laboratory. Here, we compared the results obtained from the blood serum obtained from 8810 patients using the us-cTnI and the hs-cTnI assays run in tandem on the ADVIA Centaur XP analyser. We found that in 2279 samples the concentration of cTnI measured with the ultra-sensitive method was below the detection limit, while with the high-sensitive method, only 540 were below the detection limit. We also compared results from these assays with the ultimate diagnosis of a subset of individuals. The analysis of the results below cut-off with the ultra-sensitive method showed that this method would not detect 96 cases related to heart disorder. Overall, the main finding of our research is that hs-cTnI is the preferable option and is able to be deployed effectively in the laboratory setting.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9289728PMC
http://dx.doi.org/10.1016/j.plabm.2022.e00293DOI Listing

Publication Analysis

Top Keywords

advia centaur
8
hs-ctni assays
8
ultra-sensitive method
8
detection limit
8
assays
5
comparison advia
4
ultra-sensitive
4
centaur ultra-sensitive
4
ultra-sensitive high-sensitive
4
high-sensitive assays
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!