Intraocular lenses (IOLs) from six manufacturers were measured on a Terry keratometer and on a calibrated lens bench for lens power. Differences between the two methods appeared when the IOLs were a different refractive index than polymethylmethacrylate (PMMA) and a shape other than planoconvex. These differences were consistent for ultraviolet absorbing PMMA and for meniscus lenses.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/s0886-3350(87)80081-0 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!