Background: Studies investigating changes in blood metal ion levels during the second decade of the implant lifetime in MoM hip resurfacing patients are scarce.
Methods: Patients implanted with either Birmingham Hip Resurfacing (BHR) or Articular Surface Replacement (ASR) hip resurfacings with >10 years follow-up and repeated blood metal ion measurements were identified at 2 large specialist European arthroplasty centres. After excluding patients with initial metal ion levels >7 ppb, the proportion of patients with an increase in blood metal ion levels above previously validated implant-specific thresholds (cobalt 2.15 ppb for unilateral implants, cobalt or chromium 5.5 ppb for bilateral) was assessed.
Results: We included 2743 blood metal ion measurements from 457 BHR patients (555 hips) and 216 ASR patients (263 hips). Of patients with initial metal ion levels below implant specific thresholds, increases in cobalt or chromium level, respectively, to above these thresholds during the second decade were seen as follows: unilateral BHR (cobalt = 15.6%), unilateral ASR (cobalt = 13.8%), bilateral BHR (cobalt = 8.2%, chromium = 11.8%), bilateral ASR (cobalt = 8.5%, chromium = 4.3%). Measurement-to-measurement changes exceeding +2.15 ppb or +5.5 ppb were, however, uncommon during the second decade. Subgroup results with small diameter (<50 mm) implants were similar.
Conclusions: We recommend less frequent blood metal ion measurements are needed (every 3 to 5 years) for hip resurfacing patients if initial values were below 7ppb.