Background: Numerous surgical options are available for the management of massive irreparable rotator cuff tears, but there are no current definitive guidelines concerning the optimal treatment modality. The purpose of this study was to evaluate the efficacy and safety of a biodegradable subacromial spacer (InSpace) implantation in patients with irreparable rotator cuff tears.

Methods: A retrospective study was conducted involving 47 patients treated with the InSpace balloon between 2016 and 2018. Shoulder function was assessed using Constant Score. Pain was scored using a visual analogue scale, with scores ranging from 0 to 10.

Results: At an average follow-up of 24.6 months (range 12-38), the Constant Score had improved from 39.4 to 71.5 points (p < 0.0001). The range of motion, a main component of Constant Score, was improved after two years, from 27.6 to 42.2 points (p < .001). The visual analogue scale score decreased from 6.32 at baseline to a mean score of 2.7 points (p < 0.0001).

Discussion: Arthroscopic deployment of the InSpace device was found to be a safe, reliable treatment option in patients with painful irreparable rotator cuff tears, with meaningful improvement in shoulder function without serious complications.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9284253PMC
http://dx.doi.org/10.1177/1758573220960468DOI Listing

Publication Analysis

Top Keywords

rotator cuff
12
subacromial spacer
8
cuff tears
8
retrospective study
8
irreparable rotator
8
constant score
8
spacer system
4
system irreparable
4
irreparable posterosuperior
4
posterosuperior rotator
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!