We compared ID Now™ and Hologic® Panther Aptima™ for the detection of SARS-COV-2. ID Now™ showed a positive and negative percent agreement of 86.9% and 99.7% respectively. This facilitates faster clinical decision-making, along with the rapid implementation of infection control measures, and improvement of patient flow in the emergency department toward inpatient wards.