Objective: Ultra-high-resolution CT (UHR-CT), which can be applied normal resolution (NR), high-resolution (HR), and super-high-resolution (SHR) modes, has become available as in conjunction with multi-detector CT (MDCT). Moreover, deep learning reconstruction (DLR) method, as well as filtered back projection (FBP), hybrid-type iterative reconstruction (IR), and model-based IR methods, has been clinically used. The purpose of this study was to directly compare lung CT number and airway dimension evaluation capabilities of UHR-CT using different scan modes with those of MDCT with different reconstruction methods as investigated in a lung density and airway phantom design recommended by QIBA.

Materials And Methods: Lung CT number, inner diameter (ID), inner area (IA), and wall thickness (WT) were measured, and mean differences between measured CT number, ID, IA, WT, and standard reference were compared by means of Tukey's HSD test between all UHR-CT data and MDCT reconstructed with FBP as 1.0-mm section thickness.

Results: For each reconstruction method, mean differences in lung CT numbers and all airway parameters on 0.5-mm and 1-mm section thickness CTs obtained with SHR and HR modes showed significant differences with those obtained with the NR mode on UHR-CT and MDCT (p < 0.05). Moreover, the mean differences on all UHR-CTs obtained with SHR, HR, or NR modes were significantly different from those of 1.0-mm section thickness MDCTs reconstructed with FBP (p < 0.05).

Conclusion: Scan modes and reconstruction methods used for UHR-CT were found to significantly affect lung CT number and airway dimension evaluations as did reconstruction methods used for MDCT.

Key Points: • Scan and reconstruction methods used for UHR-CT showed significantly higher CT numbers and smaller airway dimension evaluations as did those for MDCT in a QIBA phantom study (p < 0.05). • Mean differences in lung CT number for 0.25-mm, 0.5-mm, and 1.0-mm section thickness CT images obtained with SHR and HR modes were significantly larger than those for CT images at 1.0-mm section thickness obtained with MDCT and reconstructed with FBP (p < 0.05). • Mean differences in inner diameter (ID), inner area (IA), and wall thickness (WT) measured with SHR and HR modes on 0.5- and 1.0-mm section thickness CT images were significantly smaller than those obtained with NR mode on UHR-CT and MDCT (p < 0.05).

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-022-08983-1DOI Listing

Publication Analysis

Top Keywords

lung number
20
reconstruction methods
20
shr modes
20
airway dimension
16
10-mm thickness
16
number airway
12
scan modes
12
reconstructed fbp
12
reconstruction
9
dimension evaluation
8

Similar Publications

Introduction: Immunotherapy targeting PD-1/PD-L1 shows significant benefits in lung cancer. Cutaneous immune-related adverse events (irAEs) are frequent, early-developing side effects of ICIs, and their potential role as prognostic markers in non-small cell lung cancer (NSCLC) therapy requires further exploration.

Methods: Data of patients with NSCLC treated with camrelizumab Combined with chemotherapy were collected at Xuzhou Medical University from 2019 to 2023.

View Article and Find Full Text PDF

Objective: Incorporate sleep into a novel lifestyle intervention strategy in adolescents with Emerging symptoms of polycystic ovary syndrome (E-PCOS).

Design: A single-center cohort study.

Setting: University hospital-based clinic for adolescents with PCOS.

View Article and Find Full Text PDF

[Aspirin reduces lung inflammatory response in acute lung injury/acute respiratory distress syndrome: a Meta-analysis based on animal experiments].

Zhonghua Wei Zhong Bing Ji Jiu Yi Xue

December 2024

Department of Intensive Care Unit, the Affiliated Hospital of Guizhou Medical University, Guiyang 550004, Guizhou, China. Corresponding author: Shen Feng, Email:

Objective: To systematically evaluate the impact of aspirin on the pulmonary inflammatory response in animal models of acute lung injury/acute respiratory distress syndrome (ALI/ARDS).

Methods: Experimental research on aspirin therapy or prevention of ALI/ARDS in animal models were searched in PubMed, Web of Science, Cochrane library, Embase, China biology medicine, CNKI, Wanfang, VIP. The search time limit was from the establishment of the database to July 17, 2023.

View Article and Find Full Text PDF

Objective: To evaluate the efficacy and safety of magnesium sulfate in the treatment of acute severe asthma in adults.

Methods: Literature searches were conducted on PubMed, Cochrane, CNKI, VIP and Wanfang databases to screen randomized controlled trial (RCT) of magnesium sulfate in the treatment of acute severe asthma in adults, starting from the establishment of the database and ending on May 22, 2024. The control group received conventional treatment.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!