Research and development strategy for biodegradable magnesium-based vascular stents: a review.

Biomater Transl

National Engineering Research Center of Light Alloy Net Forming and Key State Laboratory of Metal Matrix Composites, School of Materials Science and Engineering, Shanghai Jiao Tong University, Shanghai, China.

Published: September 2021

Magnesium alloys are an ideal material for biodegradable vascular stents, which can be completely absorbed in the human body, and have good biosafety and mechanical properties. However, the rapid corrosion rate and excessive localized corrosion, as well as challenges in the preparation and processing of microtubes for stents, are restricting the clinical application of magnesium-based vascular stents. In the present work we will give an overview of the recent progresses on biodegradable magnesium based vascular stents including magnesium alloy design, high-precision microtubes processing, stent shape optimisation and functional coating preparation. In particular, the Triune Principle in biodegradable magnesium alloy design is proposed based on our research experience, which requires three key aspects to be considered when designing new biodegradable magnesium alloys for vascular stents application, i.e. biocompatibility and biosafety, mechanical properties, and biodegradation. This review hopes to inspire the future studies on the design and development of biodegradable magnesium alloy-based vascular stents.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9255809PMC
http://dx.doi.org/10.12336/biomatertransl.2021.03.06DOI Listing

Publication Analysis

Top Keywords

vascular stents
24
biodegradable magnesium
16
magnesium-based vascular
8
magnesium alloys
8
biosafety mechanical
8
mechanical properties
8
magnesium alloy
8
alloy design
8
stents
7
biodegradable
6

Similar Publications

Diffuse coronary artery disease (CAD) impacts the immediate hemodynamic and clinical outcomes of percutaneous coronary intervention (PCI). We evaluated whether the diffuse pattern of CAD derived from angiographic Quantitative flow ratio (QFR) impacts the immediate hemodynamic outcome post-PCI and the medium term predicted vessel-oriented composite endpoint (VOCE). Paired pre-procedure QFRs were assessed in 503 patients and 1022 vessels in the Multivessel TALENT (MVT) trial.

View Article and Find Full Text PDF

Objective: We present our experience with endovascular Bentall procedure (Endo-Bentall) using a modular valve conduit (Endo-Bentall) in high-risk patients with aortic root pathologies.

Methods: The physician constructed Endo-Bentall device is composed of a self-expanding transcatheter aortic valve (TAVR), aortic endovascular stent graft (TEVAR), and two wire-reinforced fenestrations for coronary artery stenting. The TAVR valve is sutured into an appropriately sized TEVAR graft.

View Article and Find Full Text PDF

Introduction: In 2015, Society for Vascular Surgery guidelines on claudication management were released spanning optimal medical management, procedural, and post-procedure recommendations. Uptake of guidelines and changes to clinical practice over time remain unknown. This study hypothesized that guideline aligned practice increased after guideline release.

View Article and Find Full Text PDF

Carotid artery stenting (CAS) has become a favoured alternative to surgical carotid endarterectomy (CEA) in select cases of critical internal carotid artery (ICA) occlusion. However, complications such as stent migration or entrapment can occur, necessitating prompt diagnosis and intervention. We present a case of a 75-year-old diabetic male who underwent CAS for recurrent presyncope at a private hospital, and during the procedure of CAS the stent was migrated and its proximal part was impacted in the critically narrowed part of the ICA.

View Article and Find Full Text PDF