Harmonization based on quantitative analysis of standardized uptake value variations across PET/CT scanners: a multicenter phantom study.

Nucl Med Commun

Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Published: September 2022

Objectives: This study aimed to measure standardized uptake value (SUV) variations across different PET/computed tomography (CT) scanners to harmonize quantification across systems.

Methods: We acquired images using the National Electrical Manufacturers Association International Electrotechnical Commission phantom from three PET/CT scanners operated using routine imaging protocols at each site. The SUVs of lesions were assessed in the presence of reference values by a digital reference object (DRO) and recommendations by the European Association of Nuclear Medicine (EANM/EARL) to measure inter-site variations. For harmonization, Gaussian filters with tuned full width at half maximum (FWHM) values were applied to images to minimize differences in SUVs between reference and images. Inter-site variation of SUVs was evaluated in both pre- and postharmonization situations. Test-retest analysis was also carried out to evaluate repeatability.

Results: SUVs from different scanners became significantly more consistent, and inter-site differences decreased for SUV mean , SUV max and SUV peak from 17.3, 20.7, and 15.5% to 4.8, 4.7, and 2.7%, respectively, by harmonization ( P values <0.05 for all). The values for contrast-to-noise ratio in the smallest lesion of the phantom verified preservation of image quality following harmonization (>2.8%).

Conclusions: Harmonization significantly lowered variations in SUV measurements across different PET/CT scanners, improving reproducibility while preserving image quality.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MNM.0000000000001598DOI Listing

Publication Analysis

Top Keywords

pet/ct scanners
12
standardized uptake
8
scanners
5
suv
5
harmonization
4
harmonization based
4
based quantitative
4
quantitative analysis
4
analysis standardized
4
variations
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!