Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Vital injuries during midurethral sling surgery are avoided by maintaining constant trocar contact with bone, and yet this is challenging for a teaching surgeon to monitor during this blind procedure. We modified a retropubic trocar with a load cell to distinguish on-bone and off-bone movement and tested it on a midurethral sling surgery 3-dimensional surgery simulator.
Methods: Two experts and 3 novice surgeons performed retropubic trocar passage on the physical pelvic floor model using the modified trocar. Biofidelity was assessed comparing expert performance on a Thiel-embalmed cadaver and the physical model. The test-retest was assessed comparing performance on the physical pelvic model 2 weeks apart. The force variables were analyzed with paired and independent t tests. We performed post hoc analyses comparing the experts to novices on the physical model.
Results: The root-mean-squared force was similar between the cadaver and model (24.3 vs 21.1 pounds, P = .62), suggesting biofidelity. Root-mean-squared force was also similar between the test and retest (14.0 vs 19.1 pounds, P =. 30). The expert surgeons exhibited a larger maximum force amplitude (51.2 vs 22.7 pounds, P = .03), shorter time to maximum force (2.7 vs 9.5 seconds, P = .03) and larger maximum rate of force development (171.5 vs 54.0 pounds/second, P = .01).
Conclusion: This study suggested high test-retest reliability and adequate biofidelity of the modified trocar used on our midurethral sling surgery 3-dimensional surgery simulator. This innovative trocar can be used both in simulation and in the operating room to help the novice surgeons stay on the bone and to help the attending surgeon monitor safe surgery.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9753144 | PMC |