Purpose: To evaluate the clinical utility of the Gore Excluder iliac branch endoprosthesis (IBE) for Japanese patients with aortoiliac aneurysms.

Materials And Methods: This was a multicenter retrospective cohort study (J-Preserve Registry). Patients undergoing endovascular aortic repair using the Gore Excluder IBE for aortoiliac aneurysms between August 2017 and June 2020 were enrolled. Data pertaining to the baseline and anatomical characteristics, technical details, and clinical outcomes were collected from each institution. The primary endpoints were technical success, IBE-related complications, and reinterventions. Secondary endpoints were mortality, aneurysm size change, and reintervention during follow-up. Technical success was defined as accurate deployment of the IBE without type Ib, Ic, or III endoleaks on the IBE sides on completion angiography. A change in aneurysm size of 5 mm or more was taken to be a significant change.

Results: We included 141 patients with 151 IBE implantations. Sixty-five IBE implantations (43.0%) had at least one instruction for use violation. Twenty-two patients (15.6%) required internal iliac artery (IIA) embolization for external iliac artery extension on the contralateral side. Of 151 IBE implantations, 19 exhibited IIA branch landing zones due to IIA aneurysms. Mean maximum and proximal common iliac artery (CIA) diameters were 32.9±9.9 mm and 20.5±6.9 mm, respectively. The mean CIA length was 59.1±17.1 mm. The IIA landing diameter and length were 9.0±2.3 mm and 33.8±14.6 mm. The overall technical success rate was 96.7%. There were no significant differences in IBE-related complications (2.3% vs 5.3%, =0.86) or IBE-related reinterventions (1.5% vs 5.3%, =0.33) between the IIA trunk and IIA branch landing groups. The mean follow-up period was 635±341 days. The all-cause mortality rate was 5.0%. There were no aneurysm-related deaths or ruptures during the follow-up. Most patients (95.7%) had sac stability or shrinkage.

Conclusion: The Gore Excluder IBE was safe and effective for Japanese patients in the midterm. Extending the IIA device into the distal branches of the IIA was acceptable, which may permit extending indications for endovascular aortic aneurysm repair of aortoiliac aneurysms to more complex lesions.

Clinical Impact: This study suggests clinical benefits of the Gore Excluder IBE for Japanese patients, despite 43% of the IBE implantations having at least one IFU violation.

Download full-text PDF

Source
http://dx.doi.org/10.1177/15266028221109477DOI Listing

Publication Analysis

Top Keywords

gore excluder
20
japanese patients
16
ibe implantations
16
aortoiliac aneurysms
12
excluder ibe
12
technical success
12
iliac artery
12
ibe
10
excluder iliac
8
iliac branch
8

Similar Publications

Isolated iliac aneurysms are rare, and their management becomes complex when accompanied by ectopic pelvic kidneys due to altered vascular anatomy. We report a 58-year-old male with an incidentally discovered 53.5 mm left common iliac artery (CIA) aneurysm and an ectopic pelvic kidney.

View Article and Find Full Text PDF

Objective: This study aimed to compare the completion of gate cannulation task performed by participants of varying experience using fluoroscopy, the Intraoperative Positioning System (IOPS)-a United States Food and Drug Administration-cleared endovascular navigation system that has been developed to reduce dependence on fluoroscopy-or an investigational augmented reality electromagnetic navigation technology based on IOPS.

Methods: The task consisted in the cannulation of the gate of a GORE Excluder AAA endoprosthesis bifurcated aortic stent graft (W.L.

View Article and Find Full Text PDF

Purpose: To evaluate retrospectively the 2-year outcomes of the Gore Excluder Iliac Branch Endoprosthesis (IBE) in patients with and without coexisting hypogastric artery (HA) aneurysms in a large contemporary multicentric European experience using dedicated bridging devices.

Methods: The study included all consecutive patients treated at participating institutions with the Gore Excluder IBE device who received a covered stent (i.e.

View Article and Find Full Text PDF

Objective: The physical impact of the application of Heli-FX EndoAnchors (EA; Medtronic, Minneapolis, USA) on endograft (EG) material is unclear. This study aimed to examine the possible EG membrane alterations after EA implantation.

Methods: Heli-FX EndoAnchors were applied into four aortic endocuffs: AFX2 (Endologix Inc.

View Article and Find Full Text PDF

Endovascular aneurysm repair for abdominal aortic aneurysms with a severely angulated proximal neck remains challenging because of the higher rates of type 1a endoleaks and secondary interventions. In this study, we reviewed six consecutive patients with a severely angulated proximal neck (>60°) treated with endovascular aneurysm repair at a single center. They were treated with a Gore Excluder Conformable endograft using a pre-constrain technique to prevent type 1a endoleaks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!