A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

The Togo national proficiency test pilot programme for basic clinical chemistry tests. | LitMetric

The Togo national proficiency test pilot programme for basic clinical chemistry tests.

Afr J Lab Med

Division of Laboratories - RESAOLAB, Ministry of Health and Public Hygiene, Lomé, Togo.

Published: June 2022

Background: A national proficiency test (PT) programme is not currently implemented in most low-income countries. However, participation in such PT programmes assists improves test performance and result accuracy.

Objective: This study assessed how well 11 government hospital laboratories performed 18 basic clinical chemistry tests and identified areas needing improvement.

Methods: A cross-sectional study was carried out by the Division of Laboratories of the Ministry of Health of Togo from 01 July 2016 to 31 December 2016. The test performance was evaluated using panels provided by One World Accuracy, Canada (Vancouver). The Clinical Laboratory Improvement Amendments criteria were used in evaluating the laboratories, and their success rates were compared with the World Health Organization Regional Office for Africa's target of 80%.

Results: The overall rate of acceptable results at the laboratories was over 80% for glucose, alanine aminotransferase, aspartate aminotransferase, gamma-glutamyltransferase, alkaline phosphatase and triglycerides tests. The laboratories using fully automated spectrophotometers had an acceptable results rate of 89% ( = 0.001). The overall performance of the laboratories by cycles varied from 71% to 82%.

Conclusion: This national PT programme identified the tests, which laboratories must improve their performance (urea, creatinine, uric acid, bilirubin, cholesterol, total protein, calcium, magnesium, phosphorus). It demonstrated the need for the use of routine appropriate internal quality control in all laboratories. The proficiency test programme should be extended to all clinical laboratories and target all biology disciplines.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9257935PMC
http://dx.doi.org/10.4102/ajlm.v11i1.1565DOI Listing

Publication Analysis

Top Keywords

proficiency test
12
laboratories
9
national proficiency
8
basic clinical
8
clinical chemistry
8
chemistry tests
8
test programme
8
test performance
8
tests laboratories
8
test
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!