Background: Lately, major advances in crucial aspects of magnetic hyperthermia (MH) therapy have been made (nanoparticle synthesis, biosafety, etc.). However, there is one key point still lacking improvement: the magnetic field-frequency product ( = 4.85 × 10 Ams) proposed by Atkinson-Brezovich as a limit for MH therapies. Herein, we analyze both local and systemic physiological effects of overpassing this limit.

Methods: Different combinations of field frequency and intensity exceeding the Atkinson-Brezovich limit (591-920 kHz, and 10.3-18 kA/m) have been applied for 21 min to WAG/RijHsd male rats, randomly distributed to groups of 12 animals; half of them were sacrificed after 12 h, and the others 10 days later. Biochemical serum analyses were performed to assess the general, hepatic, renal and/or pancreatic function.

Results: MH raised liver temperature to 42.8 ± 0.4 °C. Although in five of the groups the exposure was relatively well tolerated, in the two of highest frequency (928 kHz) and intensity (18 kA/m), more than 50% of the animals died. A striking elevation in liver and systemic markers was observed after 12 h in the surviving animals, independently of the frequency and intensity used. Ten days later, liver markers were almost recovered in all of the animals. However, in those groups exposed to 591 kHz and 16 kA/m, and 700 kHz and 13.7 kA/m systemic markers remained altered.

Conclusions: Exceeding the Atkinson-Brezovich limit up to 9.59 × 10 Ams seems to be safe, though further research is needed to understand the impact of intensity and/or frequency on physiological conditions following MH.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9265033PMC
http://dx.doi.org/10.3390/cancers14133084DOI Listing

Publication Analysis

Top Keywords

atkinson-brezovich limit
12
magnetic hyperthermia
8
frequency intensity
8
exceeding atkinson-brezovich
8
systemic markers
8
proposal safety
4
safety limits
4
limits vivo
4
vivo experiments
4
experiments magnetic
4

Similar Publications

Magnetic nanoparticle hyperthermia (MNH) emerges as a promising therapeutic strategy for cancer treatment, leveraging alternating magnetic fields (AMFs) to induce localized heating through magnetic nanoparticles (MNPs). However, the interaction of AMFs with biological tissues leads to non-specific heating caused by eddy currents, triggering thermoregulatory responses and complex thermal gradients throughout the body of the patient. While previous studies have implemented the Atkinson-Brezovich limit to mitigate potential harm, recent research underscores discrepancies between this threshold and clinical outcomes, necessitating a re-evaluation of this safety limit.

View Article and Find Full Text PDF

Background: Lately, major advances in crucial aspects of magnetic hyperthermia (MH) therapy have been made (nanoparticle synthesis, biosafety, etc.). However, there is one key point still lacking improvement: the magnetic field-frequency product ( = 4.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!