A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Strip detector array (SDA) for beam monitoring in radiotherapy: reconstruction of MLC parameters from multiple projections of flux. | LitMetric

Strip detector array (SDA) for beam monitoring in radiotherapy: reconstruction of MLC parameters from multiple projections of flux.

Biomed Phys Eng Express

Department of Radiation Oncology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, 02115, United States of America.

Published: July 2022

In this paper we propose and investigate a new detector with multiple strip detector arrays (SDA) for monitoring MLC shaped x-ray beams for radiotherapy treatment.Each SDA measures 1D dose profiles equivalent to dose projections. The goal of such a detector is to determine individual MLC leaf positions as well as the Monitor Units (MU) per MLC segment during radiotherapy. In the present work we investigate an optimal SDA detector configuration and reconstruction algorithm. We determine the accuracy of SDA for different treatment sites (spine, pelvis, retroperitoneum, prostate, brain SRT, SRS, lung and head and neck). We perform a simulation study accounting for different type of MLC leaf positional errors: random MLC leaf, systematic for the whole leaf bank and systematic for an individual leaf. In a similar fashion, we also account for errors in Monitor Units per segment.We demonstrate that for a broad range of IMRT treatment plans a robust reconstruction of errors is achievable with only 3 projections (3 sets of SDA oriented at at 0°, 45° and 135°). The SDA is capable of capturing both systematic errors in leaf banks and individual leaves as well as random errors sufficient for practical clinical purposes.These features of the SDA detector makes it suitable for real-time Quality Control of MLC collimated linac output.

Download full-text PDF

Source
http://dx.doi.org/10.1088/2057-1976/ac7fbcDOI Listing

Publication Analysis

Top Keywords

mlc leaf
12
strip detector
8
sda
8
monitor units
8
sda detector
8
mlc
7
leaf
6
detector
5
errors
5
detector array
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!