Introduction: A rare catastrophic failure of modular component Total Hip Arthroplasty is dissociation between liner and cup, which has been associated with component malposition and/or impingement and seems to be more frequently associated with the Pinnacle system. The goal of this study was to evaluate the resistance of a polyethylene liner to lever-out-forces of the Pinnacle locking mechanism and the locking mechanisms of two other current cup/liner systems using a standardized testing method (ASTM).
Materials And Methods: Five of each of the following cups were evaluated with their corresponding polyethylene liners: Pinnacle Multihole cup with and without intact anti-rotation tabs (ART's); Allofit-S-Alloclassic and Plasmafit Plus7 cups. The ASTM test set-up was used to evaluate the lever-out force resulting in liner dissociation for each construct.
Results: The Pinnacle construct with intact ARTs required the greatest force (F) to achieve dissociation (263.2 ± 79.2 N) followed by the Plasmafit Plus7 (185.8 ± 36.9 N) and the Allofit-S (101.4 ± 35.3 N) constructs, respectively. However, after removal of the ARTs, the Pinnacle system required the least force to achieve dissociation (75.1 ± 22.2 N) (p < 0.001).
Conclusions: The intact Pinnacle system appeared the most stable in lever-out tests when compared to the other systems. However, after removal of the ARTs, the Pinnacle system required the least force for dissociation, consistent with locking mechanism failure, and suggesting that the ARTs are a critical component of the locking mechanism. Our findings are consistent with the clinical experience of dissociated Pinnacle constructs displaying damaged or missing ARTs, and that damage to these may increase risk of liner dissociation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10110686 | PMC |
http://dx.doi.org/10.1007/s00402-022-04529-8 | DOI Listing |
Arthroplast Today
December 2024
Orthopaedic Department, University of Ulm, Ulm, Germany.
Since the early 2000s, oxidized zirconium implants have emerged as a valuable option in total hip and knee arthroplasty due to their wear resistance and suitability for patients with metal hypersensitivity. The surface of these components is created through a heating and oxidation process of a zirconium alloy, resulting in a thin layer with favorable wear properties. However, there have been few reports of severe metallosis resulting from inadvertent wear of oxidized zirconium components through various mechanisms, including dissociation of the polyethylene liner and joint instability.
View Article and Find Full Text PDFJ Shoulder Elbow Surg
October 2024
Department of Orthopaedic Surgery, Duke University, Durham, NC, USA.
Background: The purpose of this study was to compare the outcomes of primary reverse total shoulder arthroplasty (rTSA) using constrained liners (in a 145° onlay implant, Equinoxe [Exactech]) with primary rTSA using standard liners with a minimum 1-year follow-up.
Methods: A total of 836 primary rTSA patients were analyzed in this study. Patients treated with constrained liners (n = 209) were cross-matched 1:3 for age, gender, glenosphere diameter, and follow-up duration and compared with 627 patients who underwent primary rTSA with standard liners.
J Arthroplasty
September 2024
Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.
Background: Ceramic-on-highly crosslinked polyethylene (HXLPE) has become the most common bearing surface utilized in primary total hip arthroplasty (THA). The purpose of this study was to determine the implant survivorship and clinical outcomes of THAs with ceramic-on-HXLPE in a large single-institutional series.
Methods: We identified 5,536 primary THAs performed from 2007 to 2017 using a ceramic-on-HXLPE bearing through our total joint registry.
J Orthop Traumatol
August 2024
Department of Orthopaedic Surgery, University of Regensburg, Asklepios Klinikum, Kaiser-Karl-V.-Allee 3, 93077, Bad Abbach, Germany.
Background: Modular acetabular components for total hip arthroplasty (THA) provide intraoperative flexibility; however, polyethylene liner dissociation may occur. This study aimed to examine the incidence and causes of liner dissociation associated with a specific acetabular component design at a single centre.
Materials And Methods: A retrospective analysis of 7027 patients who underwent primary THA was performed to identify isolated liner dislocations.
Technol Health Care
September 2024
Hospital of Trauma Surgery, Marktredwitz Hospital, Marktredwitz, Germany.
Background: Uncemented acetabular components are widely used in modern total hip arthroplasty (THA). Modularity has numerous advantages including the ability to use supplementary screw fixation for the shell, and allow to switch from ceramic to polyethylene (PE) bearings and vice versa, and the use of lipped and face-changing liners.
Objective: Despite these advantages, a problem with modular PE liners is dissociation.