To establish a solvent desorption gas chromatographymethod for the determination of phenyl glycidyl ether (PGE) . From October to December 2020, PGE in the air of workplace was collected with carbon tube and desorbed by 25% acetone-carbon disulfide. The target toxicant was separated with the gas chromatography (GC) column and analyzed with flame ionization detector (FID), and quantified by peak area. The linear range of PGE in the air of workplace was 10.0-1109.0 μg/ml, the linear equation was =1.156-4.328, with a correlation coefficient of 0.9997. The limit of detection was 3.0 μg/ml. The lower limit of quantification was 10.0 μg/ml. The intar-batch and inter-batch precisionswere 4.9%-6.4% and 6.2%-6.9%, respectively. The recovery rate was ranged from 97.2%-98.8%, the average collection efficiency was 100%, and the average extraction efficiency was 90.1%. The samples could be stored at 4 ℃ for 7 d. This method has high precision and good accuracy, and it is applicable for the determination of PGE in workplace air.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3760/cma.j.cn121094-20210422-00228 | DOI Listing |