Background: The goal of stabilization of the femoral neck is to limit morbidity and mortality from fracture. Of three potential methods of fixation, (three percutaneous screws, the Synthes Femoral Neck System, and a dynamic hip screw), each requires guide wire positioning of the implant(s) in the femoral neck and head. Consistent and accurate positioning of these systems is paramount to reduce surgical times, stabilize fractures effectively, and reduce complications. To help expedite surgery and achieve ideal implant positioning in the geriatric population, we have developed and validated a surgical planning methodology using 3D modelling and printing technology.

Methods: Using image processing software, 3D surgical models were generated placing guide wires in a virtual model of an osteoporotic proximal femur sawbone. Three unique drill guides were created to achieve the optimal position for implant placement for each of the three different implant systems, and the guides were 3D printed. Subsequently, a trauma fellowship trained orthopedic surgeon used the 3D printed guides to position 2.8 mm diameter drill bit tipped guide wires into five osteoporotic sawbones for each of the three systems (fifteen sawbones total). Computed Tomography (CT) scans were then taken of each of the sawbones with the implants in place. 3D model renderings of the CT scans were created using image processing techniques and the displacement and angular deviations at guide wire entry to the optimal sawbone model were measured.

Results: Across all three percutaneous screw guide wires, the average displacement was 3.19 ± 0.12 mm and the average angular deviation was 4.10 ± 0.17. The Femoral Neck System guide wires had an average displacement of 1.59 ± 0.18 mm and average angular deviation of 2.81 ± 0.64. The Dynamic Hip Screw had an average displacement of 1.03 ± 0.19 mm and average angular deviation of 2.59 ± 0.39.

Conclusion: The use of custom 3D printed drill guides to assist with the positioning of guide wires proved to be accurate for each of the three types of surgical strategies. Guides which are used to place more than 1 guide wire may have lower positional accuracy, as the guide may shift during multiple wire insertions. We believe that personalized point of care drill guides provide an accurate intraoperative method for positioning implants into the femoral neck.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9254431PMC
http://dx.doi.org/10.1186/s41205-022-00146-8DOI Listing

Publication Analysis

Top Keywords

femoral neck
24
guide wires
20
guide wire
16
drill guides
16
average displacement
12
average angular
12
angular deviation
12
guide
9
accuracy guide
8
printed drill
8

Similar Publications

This case series reviews surgeries involving elderly patients with femoral neck fractures on apixaban who underwent spinal anesthesia (SA) within 72 hours of their last dose. Despite patients being on anticoagulation, no neurological complications occurred, suggesting SA may be practical in cases where the benefits of timely surgery outweigh the potential risks, including apixaban discontinuation for a period of less than the recommended 72 hours with detectable levels of the drug remaining in the plasma. Quantitative apixaban measurements offered useful anticoagulation status insights, though safe thresholds remain undefined.

View Article and Find Full Text PDF

Background: Despite multiple studies, less recent literature and data regarding the mortality associated with hip fractures in the elderly population are available.

Objectives: To assess the mortality data and functional outcomes of patients who underwent cemented and uncemented hemiarthroplasty in femoral neck fractures. To evaluate if preoperative (minimum 2 months) calcium and vitamin D supplement intake in patients affects postoperative mobilization with or without walker support.

View Article and Find Full Text PDF

Background: Hip morphology variations, particularly in femoral neck shaft angle (NSA) and iliac wing width (IWW), have been associated with gluteal tendinopathy. However, the biomechanical implications of these morphological differences on gluteal muscle function are not well understood. This study investigates how NSA and IWW influence gluteal muscle forces, moment arms, and estimated tendon loads during walking, aiming to provide insights into the potential biomechanical pathways that may contribute to altered lateral hip loading patterns.

View Article and Find Full Text PDF

Aims: The primary aim of this study is to compare mobility status of patients receiving oral oxycodone with those receiving subcutaneous alfentanil as analgesic methods prior to mobilization to help physiotherapy compliance after hip fracture surgery. The secondary aims are to assess postoperative pain, health-related quality of life, in-hospital length of stay, total use of analgesia over postoperative days 1 and 2 (POD 1 and POD 2), complication rates within 30 days, and 30-day mortality rates.

Methods: A single-centre, prospective cohort study of 64 patients will be undertaken.

View Article and Find Full Text PDF

Background: Femoroacetabular impingement (FAI) is a well-recognized cause of hip pain in adults. The hip-spine relationship between the femur, pelvis, and lumbosacral spine has garnered recent attention in hip arthroplasty. However, the hip-spine relationship has not been well described in patients with FAI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!