Evaluation of a Rapid Biological Spore Test for Dental Instrument Sterilization.

J Contemp Dent Pract

Center for Dentistry and Oral Hygiene, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.

Published: March 2022

Aim: This study evaluated the reliability of a new rapid biological spore test (BST) for determining the sterilization efficacy of dental steam autoclaves within 20 minutes, as compared to a conventional BST requiring 2 days of incubation after autoclave exposure.

Materials And Methods: A total of 177 pairs of BST, each composed of a rapid test (Celerity™ 20 Steam Biologic Indicator, Steris) and a conventional BST (Attest™ 1262 Biological Indicator, 3M), both containing spores, were placed into steam autoclaves loaded with instruments, and subjected to either sterilizing (157 pairs) or non-sterilizing conditions (20 pairs). Celerity™ BST was then incubated for 20 minutes at 57°C, with the growth medium evaluated spectrophotometrically for fluorescent α-glucosidase signal changes (no change with successful sterilization; increased fluorescence after failed sterilization). Attest™ BST was incubated for 48 hours at 57°C, after which a pH-based color change in the culture broth was visually assessed (no change in purple color with successful sterilization; change to yellow color with failed sterilization).

Results: Celerity™ and Attest™ BST both accurately identified successful sterilization, with no spore growth from either BST after exposure to sterilizing steam autoclave conditions (100% agreement between 157 pairs of each BST). Both BST also accurately detected unsuccessful sterilization, with all tested ampoules positive for spore germination after non-sterilizing steam autoclave time periods. Both BST exhibited 100% sensitivity, specificity, and accuracy for detection of sterilizing steam autoclave conditions.

Conclusion: Celerity™ BST, after only 20 minutes incubation, performed equally as well as a BST requiring 48 hours incubation in determining the sterilization efficacy of dental steam autoclaves.

Clinical Significance: Rapid BST offer earlier detection of sterilization failure before potentially contaminated dental instruments are used in clinical patient care.

Download full-text PDF

Source

Publication Analysis

Top Keywords

bst
14
successful sterilization
12
steam autoclave
12
sterilization
9
rapid biological
8
biological spore
8
spore test
8
determining sterilization
8
sterilization efficacy
8
efficacy dental
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!