Prcis: Corneal hysteresis (CH) measurements are relatively stable during routine follow up of glaucoma patients over an intermediate time frame. Cataract surgery does not change the CH significantly after an average follow up of 6 months.

Purpose: The aim was to assess the repeatability of CH measurements in glaucoma patients over time, during routine follow up and after cataract surgery.

Materials And Methods: Retrospective analysis of patients records in a glaucoma clinic where routine measurements by the Ocular Response Analyzer were done. Patients with at least 2 CH measurements were included. Repeatability was assessed using the intraclass correlation coefficient (ICC).

Results: A total of 164 eyes (87 patients) were included. Twenty-eight of them had measurements before and after cataract surgery with mean follow up time of 29.64±9.63 weeks. There was no evidence for a difference in CH between the before and after cataract surgery measurements (ICC=0.79, 95% confidence interval: 0.60-0.89). In the remaining 136 eyes, without any surgical treatment between measurements, there was moderate agreement among the repeated CH measurements (ICC=0.64, 95% confidence interval: 0.61-0.82) with mean time of 32.06±25.32 weeks between first and last measurement.

Conclusion: CH measurements in glaucoma patients were repeatable over a 6-month period during routine follow up or following cataract surgery. These findings suggest that the intraocular pressure reduction following cataract surgery is unlikely to be because of a change in this biomechanical property.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9240401PMC
http://dx.doi.org/10.1097/IJG.0000000000002008DOI Listing

Publication Analysis

Top Keywords

cataract surgery
24
glaucoma patients
16
routine follow
16
measurements glaucoma
12
follow cataract
12
measurements
10
corneal hysteresis
8
hysteresis measurements
8
surgery change
8
95% confidence
8

Similar Publications

Background: This study aimed to evaluate mydriasis stability during cataract surgery in patients with systemic comorbidities such as diabetes mellitus (DM) and pseudoexfoliation syndrome (PXF) after a standardised combination of intracameral mydriatics and anaesthetic (SCIMA). Stable mydriasis is crucial for safe and effective phacoemulsification.

Methods: Patients were included if they achieved pupil dilation ≥6.

View Article and Find Full Text PDF

Background: The aim of this study was to evaluate visual outcomes and patient satisfaction after bilateral implantation of a new hydrophobic acrylic intraocular lens called Clareon (Alcon) using the mini-monovision technique.

Methods: A single-center, prospective, nonrandomized study was conducted in Tandil (Buenos Aires, Argentina), including patients scheduled for cataract surgery. To achieve mini-monovision, the spherical equivalent was calculated between -0.

View Article and Find Full Text PDF

To compare the accuracy of seven artificial intelligence (AI)-based intraocular lens (IOL) power calculation formulas in medium-long Caucasian eyes regarding the root-mean-square absolute error (RMSAE), the median absolute error (MedAE) and the percentage of eyes with a prediction error (PE) within ±0.5 D. Data on Caucasian patients who underwent uneventful phacoemulsification between May 2018 and September 2023 in MW-Med Eye Center, Krakow, Poland and Kyiv Clinical Ophthalmology Hospital Eye Microsurgery Center, Kyiv, Ukraine were reviewed.

View Article and Find Full Text PDF

Waste and Greenhouse Gas Emissions Produced from Ophthalmic Surgeries: A Scoping Review.

Int J Environ Res Public Health

December 2024

Department of Ophthalmology & Visual Sciences, Montefiore Medical Center, Albert Einstein College of Medicine, New York, NY 10461, USA.

(1) Background: Healthcare is a major contributor to global greenhouse gas (GHG) emissions, especially within the surgical suite. Ophthalmologists play a role, since they frequently perform high-volume procedures, such as cataract surgery. This review aims to summarize the current literature on surgical waste and GHG emissions in ophthalmology and proposes a framework to standardize future studies.

View Article and Find Full Text PDF

Background: Endothelin is a potent vasoconstrictor and contributes to the regulation of vascular perfusion. Aberrant endothelin-1 (ET-1) levels in aqueous humor have been reported across a variety of vascular diseases of the eye, including glaucoma. These findings suggest that dysregulation of ET-1 production may contribute to glaucoma pathophysiology.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!