Background: Rapid diagnosis of infectious pathogens at an early stage is crucial to stabilize the patient's condition, reduce medical costs, and shorten hospital stays. Currently, some point-of-care tests have their own shortcomings. Therefore, we built a microfluidic chip based on loop-mediated isothermal amplification to can quickly and sensitively detect infectious pathogens.

Methods: We extracted the DNA of S. aureus, MRSA, Shigella and Klebsiella pneumoniae. Then, the DNA samples were diluted by 10-fold and examined by two methods: LAMP-microfluidic chip and q-PCR, the sensitivity of whom was also compared. In addition, the specificity of the two was also examined by detecting the target bacteria and other microorganisms using the same methods. Finally, we extracted and tested the DNA of clinically infected humoral samples to determine the coincidence rate between the two methods and the bacterial culture method.

Results: For S. aureus, MRSA, Shigella, and Klebsiella pneumoniae, the detection limits of the chip were 2.25 × 10 copies/μl, 5.32 × 10 copies/μl, 2.89 × 10 copies/μl, 6.53 × 10 copies/μl, and the detection limits of q-PCR were 2.25 × 10 copies/μl, 5.32 × 10 copies/μl, 2.89 × 10 copies/μl, 6.53 × 10 copies/μl, respectively. In terms of detection specificity, neither method cross-reacted with other strains. For the detection of infectious humoral samples, the total coincidence rate between the q-PCR and bacterial culture method was 85.7%, 95%, 95%, and 95.5%, and the total coincidence rate between the chip and bacterial culture method was 81%, 95%, 90%, and 86.4%, respectively.

Conclusion: LAMP-microfluidic chip provides a simple, sensitive, specific, convenient, and rapid pathogen detection method for clinically infected humoral samples without relying on expensive equipment or technical personnels.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9396168PMC
http://dx.doi.org/10.1002/jcla.24565DOI Listing

Publication Analysis

Top Keywords

humoral samples
12
coincidence rate
12
bacterial culture
12
loop-mediated isothermal
8
isothermal amplification
8
microfluidic chip
8
infectious pathogens
8
s aureus mrsa
8
mrsa shigella
8
shigella klebsiella
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!