  • The study evaluates the performance of two antimicrobial susceptibility testing systems, VITEK2 and Sensititre, in detecting colistin resistance in clinical isolates.
  • A total of 213 clinical isolates were tested, revealing that Sensititre showed better agreement with manual reference methods (99.1%) compared to VITEK2 (95.3%), which had a high rate of very major errors (9.9%).
  • Overall, both automated systems performed adequately for detecting colistin resistance, but challenges remain, especially with VITEK2's error rate and Sensititre's inconsistent minimum inhibitory concentration (MIC) results.*

Article Abstract

Performances of the colistin antimicrobial susceptibility testing (AST) systems of vary depending on the manufacturer, and data on colistin-resistant are limited. We evaluated the VITEK2 and Sensititre systems to determine colistin resistance and minimum inhibitory concentration (MIC) for isolated from a clinical microbiology laboratory. A total of 213 clinical isolates were tested, including 81 colistin-resistant . ASTs were performed using the VITEK2 and Sensititre systems according to the manufacturer's instructions. Reference MICs for colistin were determined using the manual broth microdilution method (BMD). The results of the two AST methods were compared with the BMD results. VITEK2 and Sensititre systems showed category agreements of 95.3% and 99.1%, respectively. VITEK2 had a relatively high very major error (VME) rate (9.9%). Sensititre reported higher MICs than the reference method for the susceptible isolates and showed low essential agreement. In conclusion, the automated systems investigated in this study showed good category agreements for colistin AST of . However, VITEK2 had a high VME rate, and Sensititre had differences in MIC results. Colistin AST remains a challenging task in the clinical laboratory.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9221671PMC
http://dx.doi.org/10.3390/diagnostics12061487DOI Listing

Publication Analysis

Top Keywords

vitek2 sensititre
16
sensititre systems
16
systems determine
8
determine colistin
8
colistin resistance
8
category agreements
8
vitek2 high
8
vme rate
8
colistin ast
8
vitek2
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!