Objective: To evaluate effectiveness of self-designed adjustable cannulated screw guide, and to provide an effective auxiliary tool for inverted triangular arrangement of compression cannulated screws in clinical treatment for transcervical femoral neck fractures.
Methods: The sketch of instrument was drawn with Solidworks software, and physical product was obtained after production. The data were obtained by Mimics software. Combined with the guide, it was first used on 22 cadaveric bones, 22 dry cadaveric bones, including 12 males and 10 females. Then the distribution of guide pins was evaluated by X-ray film. The anatomical size and screw distance of femoral head and neck were measured in different ways, and statistically compared. From January 2018 to June 2020, 45 hospitalized patients with femoral neck fracture were selected and divided into new guide group (22 patients) and free hand nail group (23 patients) according to whether the instrument was used or not. The clinical data and operation conditions between two groups were recorded and compared.
Results: The anatomical data of X-ray, three-dimensional and physical measurement were basically the same, whlie had no difference (>0.05). There was no significant difference between physical measurement and three-dimensional measurement (>0.05). The distance between screws and needle entry point was designed as an isosceles triangle(=0.992 8, <0.000 1), but due to the existence of femoral anteversion and torsion angle, it was an approximate isosceles triangle in the femoral neck (=0.824 1, <0.000 1). The patients between two groups were followed up for an average of 2 years. There was no significant difference in the number of fluoroscopy and puncture between new guide group and free hand nail group(>0.05). The screw parallelism was better and operation time was shorter which had statistically difference(<0.05). However, there was no significant difference in final Harris score and incidence of complications between two groups(>0.05).
Conclusion: Self-made femoral neck cannulated screw guide combined with preoperative planning of Mimics software is conducive to placement of inverted triangular arrangement of cannulated screws, but it still needs to be improved and followed up in the later large-scale use.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.12200/j.issn.1003-0034.2022.06.004 | DOI Listing |
PLoS One
January 2025
School of Clinical Medicine, Guizhou Medical University, Guiyang, China.
Legg-Calvé-Perthes disease (LCPD) involves femoral head osteonecrosis caused by disrupted blood supply, leading to joint deformity and early osteoarthritis. This study investigates the role of miRNA-223-5p in regulating hypoxia-induced apoptosis and enhancing osteogenesis in bone marrow mesenchymal stem cells (BMSCs). Utilizing a juvenile New Zealand white rabbit model of LCPD established through femoral neck ligation, we transfected BMSCs with miR-223-5p mimics, inhibitors, and controls, followed by hypoxic exposure.
View Article and Find Full Text PDFJ Anat
January 2025
Department of Orthopedics, Affiliated Zhongshan Hospital of Dalian University, Dalian, China.
The primary weight-bearing structure of the proximal femur, trabecular bone, has a complex three-dimensional architecture that was previously difficult to comprehensively display. This study examined the spatial architecture of trabecular struts in the coronal, sagittal, and horizontal sections of the proximal femur using 21 cases prepared with P45 sectional plasticization. The primary compressive strut (PCS) exhibited a "mushroom-like" shape with upper and lower parts.
View Article and Find Full Text PDFJ Pers Med
January 2025
Unit of Pediatric Orthopedics and Traumatology, IRCCS Istituto Ortopedico Rizzoli, 40136 Bologna, Italy.
: Femoral neck fractures are rare but serious injuries in children and adolescents, often resulting from high-energy trauma and prone to complications like avascular necrosis (AVN) and nonunion. Even rarer is the development of slipped capital femoral epiphysis (SCFE) following femoral neck fracture, which presents unique diagnostic and treatment challenges. SCFE can destabilize the femoral head, with severe cases requiring complex surgical interventions.
View Article and Find Full Text PDFThe objectives were to evaluate the effectiveness and safety of a single preoperative dose of intravenous tranexamic acid (TXA) in reducing perioperative blood loss and requirement for transfusion in patients undergoing hip hemiarthroplasty for femoral neck fracture. A double-blind randomized controlled trial was conducted in 140 patients with hip fracture. After randomization, 68 patients received a single dose of 1 gr of intravenous TXA at the start of the surgery (TXA group), and 72 received a placebo treatment (placebo group).
View Article and Find Full Text PDFProspective study of femoral neck system (FNS) vs. cannulated compression screw (CCS) fixation has not been appropriately reported. We prospectively investigate the efficacy of FNS vs.