Introduction: Titanium dental implants has been coated with different materials such as polymers and biomimetic agents, bone morphogenetic protein, calcium phosphate to enhance surface properties of the titanium implants for osseointegration. The aim of this study was to evaluate the bone tissue healing around Boron Nitride-coated (BN-coated) titanium implants histomorphometrically and biomechanically and also observe the effect of different coating thicknesses on osseointegration.
Materials And Methods: BN was coated on dental titanium implants with two different coating thicknesses by using RF magnetron sputtering system. Totally fifty-four implants were inserted into the tibias' of 12 New Zealand rabbits bilaterally under general anesthesia. All animals were sacrificed after 4-weeks. Bone-implant contact (BIC) and new bone area/total area ratios (BATA) were calculated. Also, the removal torque (RT) test was performed.
Results: The highest new bone area in the medullary cavity was around the nano-BN-coated surface with 15.70%. In micro-BN-coated surface and control group, this ratio was determined as 10.48% and 8.23%, respectively. The BIC ratios in upper-side of implants and cortical-associated BIC ratios in lower-side were found significantly higher in control and micro-BN-coated group than nano-BN-coated group (p < 0.05). Similar BIC values were observed between control and micro-BN-coated groups (p > 0.05). BATA values did not show statistically significant differences between all three groups (p > 0.05). The RT values measured in all groups were found comparable and no statistically significant differences were found (p > 0.05).
Conclusion: No inflammatory reaction developed around any implant. Relatively more new bone formation around nano-BN-coated titanium implants indicates the promising osseoinductive effect of BN coating. BN-coated implants showed similar biomechanical and histomorphometrical outcomes to that of the conventional titanium implants through a 4-week evaluation period.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jormas.2022.06.016 | DOI Listing |
Dalton Trans
January 2025
The Department of Chemistry, Karadeniz Technical University, 61080, Trabzon, Turkey.
The utilisation of implantable medical devices has become safer and more prevalent since the establishment of sterilisation methods and techniques a century ago. Nevertheless, device-associated infections remain a significant and growing concern, particularly in light of the continued rise in the number of medical device implantations. This underscores the imperative for the development of efficacious prevention and treatment strategies for device-associated infections, as well as further investigation into the design of innovative antibacterial surfaces for medical device applications.
View Article and Find Full Text PDFBiomater Sci
January 2025
Biotechnology Centre, The Silesian University of Technology, B. Krzywoustego 8, 44-100, Gliwice, Poland.
Metallic biomaterials are extensively used in orthopedics and dentistry, either as implants or coatings. In both cases, metal ions come into contact with surrounding tissues causing a particular cell response. Here, we present a biofabricated tissue model, consisting of a hydrogel reinforced with a melt electrowritten mesh, to study the effects of bound and released metal ions on surrounding cells embedded in a hydrogel matrix.
View Article and Find Full Text PDFBMC Oral Health
January 2025
Department of Prosthodontics, Faculty of Dentistry, Hacettepe University, 06230, Sıhhiye/Altındağ, Ankara, Turkey.
Background: The mechanical properties of framework materials significantly influence stress distribution and the long-term success of implant-supported prostheses. Although titanium, cobalt-chromium, zirconia, and polyether ether ketone (PEEK) are widely used, their biomechanical performance under dynamic loading conditions remains insufficiently investigated. This study aimed to evaluate the biomechanical behavior of four framework materials with different Young's modulus using dynamic finite element stress analysis.
View Article and Find Full Text PDFInt J Biol Macromol
January 2025
Departamento de Química Orgánica y Farmacéutica, Facultad de Farmacia, Universidad de Sevilla, Seville, Spain. Electronic address:
Titanium is widely used for implants however it presents limitations such as infection risk, stress shielding phenomenon, and poor osseointegration. To address these issues, a novel approach was proposed that involves fabricating porous titanium substrates, to reduce implant stiffness, minimizing stress shielding and bone resorption, and applying polymeric coatings to improve bioactivity. Composite coating prepared from chitosan, silver nanoparticles, and nanohydroxyapatite was optimized to enhance antibacterial properties and promote osseointegration.
View Article and Find Full Text PDFJ Mech Behav Biomed Mater
January 2025
Center of Stomatological, The Second Hospital, Cheeloo College of Medicine, Shandong University, 250033, PR China; Jinan Key Laboratory of Oral Materials, Jinan, 250033, PR China. Electronic address:
A method is proposed for 3D printing and enhancing the surface bioactivity of zirconia ceramic anchor screws, specifically tailored for temporomandibular joint disc reduction surgery. Initially, the challenge posed by the brittleness and processing difficulties of fine ceramic anchor screws was addressed through the application of SLA-3D printing technology. This allowed for an exploration of the forming accuracy and biomechanical properties of the printed anchor screws.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!