Background: Therapeutic drug monitoring (TDM) is useful to assess clozapine adherence and optimize treatment. However, analysis of venous blood levels by liquid chromatography tandem mass spectrometry (LC-MS/MS) is often logistically complicated and process time is prolonged.

Objective: To assess the feasibility and reliability of a new point-of-care device, (MyCare™ Insite), using capillary blood for clozapine therapeutic monitoring.

Methods: Matched venous and capillary blood samples were collected from patients treated with clozapine on a stable dose. Samples were analyzed by LC-MS/MS and MyCare Insite Clozapine Test. Clozapine plasma levels were compared between methods using linear regression model. Both patients and treatment team completed questionnaires about the feasibility of blood sampling.

Results: Of the total sample (44 patients, 61% males, mean age 43 ± 12 years), mean daily clozapine dose was 293 ± 134 mg/day. Linear regression model demonstrated high correlation with  = 0.83 ( < 0.0001) and mean difference of 26 ± 162 ng/ml. More than 60% of the patients found the clozapine TDM to be important. Most of the participants (58%) favored the capillary sampling and 11% claimed that testing method would affect their adherence to TDM. Moreover, a larger portion (72%) strongly preferred to be tested at the office rather than at the lab.

Conclusions: The point-of-care device offers an accessible and satisfactory measurement of clozapine blood levels. Both patients and healthcare providers reported preference for capillary sampling as well as for the in-office TDM procedure. The immediate results provided by the device can facilitate rapid and informed clinical decisions and therefore improve clozapine treatment outcomes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9201354PMC
http://dx.doi.org/10.1177/20451253221094435DOI Listing

Publication Analysis

Top Keywords

point-of-care device
8
feasibility reliability
8
capillary blood
8
linear regression
8
regression model
8
clozapine
7
clozapine blood
4
blood level
4
level assessment
4
assessment point-of-care
4

Similar Publications

Trigger valves are fundamental features in capillary-driven microfluidic systems that stop fluid at an abrupt geometric expansion and release fluid when there is flow in an orthogonal channel connected to the valve. The concept was originally demonstrated in closed-channel capillary circuits. We show here that trigger valves can be successfully implemented in open channels.

View Article and Find Full Text PDF

Rapid optical determination of salivary cortisol responses in individuals undergoing physiological and psychological stress.

Sci Rep

December 2024

Research Centre for Biomedical Engineering (RCBE), School of Science and Technology, City, University of London, Northampton Square, London, EC1V 0HB, UK.

Traditional methods for management of mental illnesses in the post-pandemic setting can be inaccessible for many individuals due to a multitude of reasons, including financial stresses and anxieties surrounding face-to-face interventions. The use of a point-of-care tool for self-management of stress levels and mental health status is the natural trajectory towards creating solutions for one of the primary contributors to the global burden of disease. Notably, cortisol is the main stress hormone and a key logical indicator of hypothalamic-pituitary adrenal (HPA) axis activity that governs the activation of the human stress system.

View Article and Find Full Text PDF

Advancing microfluidic point-of-care platelet function tests: opportunities and challenges from bench to market.

Front Bioeng Biotechnol

December 2024

George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA, United States.

Platelets are critical for blood clotting, with shear-induced platelet aggregation (SIPA) playing a key role in hemostasis and the prevention of excessive bleeding. SIPA function potentially leads to life-threatening diseases such as hemorrhage and myocardial infarction, which are leading causes of death globally. Point-of-care platelet function tests (POC PFTs) are developed to assess platelet dysfunction and distinguish between normal and abnormal platelet activity.

View Article and Find Full Text PDF

Background: Atrial fibrillation (AF) is the most common cardiac arrhythmia. Given its often-paroxysmal nature, screening at a single time point, using a 12-lead electrocardiogram (ECG) or a Holter monitor, has limited benefit. The AliveCor KardiaMobile device is a validated ECG recorder that can be used for patient-directed arrhythmia diagnosis and symptom-rhythm correlation.

View Article and Find Full Text PDF

Biomimetic fluorescence-enhanced platform based on photonic crystals and DNAzyme walker for visualization and quantification of miRNA-21.

Talanta

December 2024

Institute of Biomedical Precision Testing and Instrumentation, College of Artificial Intelligence, Taiyuan University of Technology, Taiyuan, Shanxi, 030600, PR China.

Developing a fluorescence sensing platform for point-of-care detection of low abundance biomarkers is highly valuable for early diagnosis of disease. Herein, a biomimetic fluorescence-enhanced platform based on photonic crystals and DNAzyme walker was constructed and further applied to visualize and quantify the miRNA-21 in biological samples. The DNAzyme walker was orthogonally activated by the target miRNA-21, which enabled the unlocking of the DNAzyme walker strand and the subsequently repeated substrate cleavage, thus generating enhanced fluorescence signals.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!