Objective: To report the personalized decision-making pro- cess adopted for a cochlear implant (CI) candidate requiring magnetic resonance imaging (MRI) brain surveillance.

Study Design: Clinical capsule report.

Setting: Tertiary academic referral center.

Patient: A 23-year-old man affected by posttraumatic bilat- eral profound hearing loss, already in radiological follow-up for a suspected small left cuneal low-grade glioma.

Interventions: A multidisciplinary approach involving preoperative MRI simulations and 3D printed (3DP) models aiming to adapt the CI position to facilitate MRI brain lesion visibility.

Main Outcome Measures: MRI visibility and surgical approach.

Results: Preoperative MRI scans with the placement of an Ultra 3D CI were performed simulating different implant location to assess the brain lesion visibility in MRI. CI was positioned 9 cm away from the external auditory canal with an angle of 90 degrees. To assess the technical feasibility of the surgical procedure, a patient-specific 3DP head model was produced preoperatively. The postoperative course was uneventful, the patient showed a significant benefit from CI, and the brain lesion was highly visible at the MRI follow-up.

Conclusions: The employment of strategies aimed at improving the MRI quality in CI recipients still represents a topic requiring attention. Thanks to multidisciplinary team collaboration, in our case, the CI position was successfully determined to allow unhindered MRI visibility of a specific intracranial structure.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAO.0000000000003532DOI Listing

Publication Analysis

Top Keywords

brain lesion
12
mri
10
cochlear implant
8
mri brain
8
preoperative mri
8
mri visibility
8
customized cochlear
4
implant positioning
4
positioning patient
4
patient low-
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!